FDA Adverse Event Injury Summary report: N

MINIMED QUICK-SET

MDR report key: 21802776 · Received April 8, 2025

Report

Report Number
3003442380-2025-05735
Event Type
Injury
Date Received
April 8, 2025
Date of Event
March 10, 2025
Report Date
July 5, 2025
Manufacturer
UNOMEDICAL DEVICES S.A. DE C.V.
Product Code
FPA
PMA / PMN Number
K160648
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

E1: PATIENT CITY - (B)(6). PATIENT COUNTRY - UNITED ARAB EMIRATES.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS - H11: INVESTIGATION SUMMARY: THE INFORMATION IN THIS COMPLAINT (B)(4) HAS BEEN EVALUATED FOR THE MALFUNCTION CODE NO MALFUNCTION BASED ON COMPLAINT INFORMATION. THE BATCH 6001589 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. COMPLAINT INVESTIGATION: THE REFERENCE SAMPLES CANNOT BE TESTED BECAUSE THERE WAS NO SPECIFIC MALFUNCTION TO INVESTIGATE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6001589 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 75, PACKAGED IN THE MULTIVAC 12, ON 31/MAY/2023, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WERE IDENTIFIED RELATED TO THE MALFUNCTION REPORTED, NO MAINTENANCE EVENTS WERE RECORDED. TRENDING: A QUERY WAS RUN IN DATABASE ON 04/JUL/2025 AGAINST MALFUNCTION CODE NO MALFUNCTION BASED ON COMPLAINT INFORMATION, HARM CODE UNTREATED DIABETIC KETOACIDOSIS WHICH THE PATIENT IS UNABLE TO SELF-MANAGE REQUIRING INTERVENTION BY AN HEALTH CARE PROFESSIONAL OR REQUIRES EMERGENCY ADVANCED LIFE SUPPORT TO PREVENT PERMANENT ORGAN DAMAGE (ELEVATED BLOOD GLUCOSE LEVEL AND SYMPTOMS E.G., MODERATE TO LARGE KETONES, NAUSEA, VOMITING, ABDOMINAL PAIN, CONFUSION) AND LOT 6001589 AND NO OTHER COMPLAINTS HAVE BEEN REGISTERED IN DATABASE FOR THE SAME LOT NUMBER, HARM CODE AND MALFUNCTION CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO OTHER COMPLAINT RECEIVED ON THE LOT IN QUESTION, HARM CODE AND MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION OR CAPA PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED ARAB EMIRATES. IT WAS REPORTED THAT THE PATIENT EXPERIENCED UNWELL, HEADACHE AND WENT TO THE EMERGENCY ROOM (ER) ON (B)(6) 2025 DUE TO HYPERGLYCEMIA. THE PATIENT WAS ADMITTED FOR LESS THAN TWENTY FOUR HOURS. THE BLOOD GLUCOSE LEVEL WAS 600 MG/DL AT THE TIME OF EVENT AND THE PATIENT GOT TREATED WITH INSULIN PEN. NO FURTHER INFORMATION AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
472258 MINIMED QUICK-SET UNO QUICK-SET 110/9 SC1 MECA FPA UNOMEDICAL DEVICES S.A. DE C.V. MMT-396A 6001589

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention