FISHER & PAYKEL HEALTHCARE
Report
- Report Number
- 9611451-2025-00281
- Event Type
- Malfunction
- Date Received
- April 8, 2025
- Report Date
- March 13, 2025
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- CAF
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- 003
Narratives
(B)(6). D2, G4: 900PT562 IS NOT SOLD IN THE USA, BUT IT IS SIMILAR TO 900PT563 WHICH IS SOLD IN THE USA. THE 510(K) FOR THE SIMILAR PRODUCT IS K211560. D4, H4: FISHER & PAYKEL HEALTHCARE (F&P) HAVE REQUESTED FOR COMPLETE DEVICE IDENTIFICATION INFORMATION, HOWEVER, THIS HAS NOT BEEN PROVIDED. FISHER & PAYKEL HEALTHCARE (F&P) HAS REQUESTED FOR THE COMPLAINT DEVICE TO BE RETURNED FOR EVALUATION. WE WILL PROVIDE A FOLLOW UP REPORT UPON COMPLETION OF OUR INVESTIGATION. PRODUCT BACKGROUND: THE HEATED BREATHING TUBE (HBT) IS A COMPONENT DESIGNED FOR USE WITH THE AIRVO 2 HUMIDIFIER (AIRVO 2), FOR THE DELIVERY OF HUMIDIFIED RESPIRATORY GASES TO PATIENTS, INCLUDING THOSE WHO ARE RECEIVING NASAL HIGH FLOW (NHF) THERAPY. NHF THERAPY IS INTENDED FOR USE WITH SPONTANEOUSLY BREATHING PATIENTS WHO WOULD BENEFIT FROM RECEIVING HIGH FLOW WARMED AND HUMIDIFIED RESPIRATORY GASES. THE AIRVO 2 DEVICE SHOULD NOT BE USED FOR LIFE SUPPORT PURPOSES, AND APPROPRIATE PATIENT MONITORING MUST BE USED AT ALL TIMES. THE HBT AS PART OF THE AIRVO 2 SYSTEM CONTAINS A HEATER WIRE ENCAPSULATED IN PLASTIC WHICH ENSURES OPTIMAL TEMPERATURE AND HUMIDIFICATION LEVELS ARE DELIVERED TO THE PATIENT INTERFACE WHILE MINIMIZING THE AMOUNT OF CONDENSATE IN THE TUBE.
(B)(6). CORRECTED DATA: B5, D9, G3, G6, H2, H3, H6. D2, G4: 900PT562 IS NOT SOLD IN THE USA, BUT IT IS SIMILAR TO 900PT563 WHICH IS SOLD IN THE USA. THE 510(K) FOR THE SIMILAR PRODUCT IS K211560. D4, H4: FISHER & PAYKEL HEALTHCARE (F&P) HAVE REQUESTED FOR COMPLETE DEVICE IDENTIFICATION INFORMATION; HOWEVER, THIS HAS NOT BEEN PROVIDED. PRODUCT BACKGROUND: THE HEATED BREATHING TUBE (HBT) IS A COMPONENT DESIGNED FOR USE WITH THE AIRVO 2 HUMIDIFIER (AIRVO 2), FOR THE DELIVERY OF HUMIDIFIED RESPIRATORY GASES TO PATIENTS, INCLUDING THOSE WHO ARE RECEIVING NASAL HIGH FLOW (NHF) THERAPY. NHF THERAPY IS INTENDED FOR USE WITH SPONTANEOUSLY BREATHING PATIENTS WHO WOULD BENEFIT FROM RECEIVING HIGH FLOW WARMED AND HUMIDIFIED RESPIRATORY GASES. THE AIRVO 2 DEVICE SHOULD NOT BE USED FOR LIFE SUPPORT PURPOSES, AND APPROPRIATE PATIENT MONITORING MUST BE USED AT ALL TIMES. THE HBT AS PART OF THE AIRVO 2 SYSTEM CONTAINS A HEATER WIRE ENCAPSULATED IN PLASTIC WHICH ENSURES OPTIMAL TEMPERATURE AND HUMIDIFICATION LEVELS ARE DELIVERED TO THE PATIENT INTERFACE WHILE MINIMISING THE AMOUNT OF CONDENSATE IN THE TUBE. METHOD: THE SUBJECT HBT AS PART OF THE 900PT562 AIRVO TUBE AND CHAMBER KIT WITH NEBULISER ADAPTOR WAS RECEIVED AT FISHER & PAYKEL HEALTHCARE (F&P) IN NEW ZEALAND FOR INVESTIGATION, WHERE IT WAS RESISTANCE TESTED AND VISUALLY INSPECTED. THE HEALTHCARE FACILITY ALSO PROVIDED ADDITIONAL INFORMATION RELATING TO THE REPORTED EVENT UPON REQUEST. F&P'S INVESTIGATION IS THUS BASED ON THE EVALUATION OF THE SUBJECT HBT, THE INFORMATION PROVIDED BY THE CUSTOMER AND F&P'S KNOWLEDGE OF THE PRODUCT. RESULTS: THE RESISTANCE CHECK ON THE HEATER WIRE WAS WITHIN SPECIFICATION. THE VISUAL INSPECTION REVEALED THAT THE HBT WAS MELTED AND FLATTENED IN THE MIDDLE OF THE TUBING, EXPOSING THE HEATER WIRES. THE PHOTOGRAPH PROVIDED BY THE HEALTHCARE FACILITY DISPLAYED THE SUBJECT DEVICE PARTIALLY COVERED AND POSITIONED ACROSS THE PATIENT'S BED. THE HEALTHCARE FACILITY STATED THAT THERE WAS NO PATIENT INJURY. CONCLUSION: F&P IS UNABLE TO DETERMINE THE EXACT CAUSE OF THE REPORTED EVENT. HOWEVER, DAMAGE TO THE HBT MAY BE CAUSED BY FACTORS SUCH AS INCORRECT SET-UP OR PARTIALLY COVERING IT WITH A MATERIAL OR OBJECT FOR A PROLONGED PERIOD. THE 900PT562 USER INSTRUCTIONS SHOW IN PICTORIAL FORMAT THE CORRECT PLACEMENT OF THE DEVICE AND INCLUDES THE FOLLOWING INFORMATION: - "CONNECT BREATHING TUBE CLIP TO PATIENT CLOTHING OR BEDDING." - "NEVER OPERATE THE UNIT IF THE BREATHING TUBE HAS BEEN DAMAGED WITH HOLES, TEARS OR KINKS". - "DO NOT BLOCK THE FLOW OF AIR THROUGH THE UNIT AND BREATHING TUBE." - "DO NOT ADD HEAT TO ANY PART OF THE BREATHING TUBE E.G., COVERING WITH A BLANKET, OR HEATING IT IN AN INCUBATOR OR OVERHEAD HEATER FOR A NEONATE, AS THIS COULD RESULT IN SERIOUS INJURY." - "DO NOT ALLOW THE BREATHING TUBE TO REMAIN IN DIRECT CONTACT WITH SKIN FOR PROLONGED PERIODS OF TIME." ALL HBTS AS PART OF THE 900PT562 AIRVO TUBE AND CHAMBER KIT ARE VISUALLY INSPECTED AND UNDERGO FUNCTIONAL TESTS, INCLUDING SOAK AND TEMPERATURE, AND HEATER WIRE RESISTANCE. THE HEATED BREATHING TUBES ARE 100% VISUALLY INSPECTED USING A CAMERA SYSTEM. THE HBTS ARE ALSO TESTED FOR RESISTANCE, CONTINUITY, POLARITY AND PITCH DURING PRODUCTION. ADDITIONALLY, A FUNCTIONAL TEST IS CONDUCTED UNDER LOAD. THE HBT WOULD HAVE MET THE REQUIRED SPECIFICATIONS AT THE TIME OF PRODUCTION. THE AIRVO 2 SYSTEM IS DESIGNED TO COMPLY WITH THE ELECTRICAL SAFETY STANDARD IEC 60601-1: 2005+A1:2012. THE CASE IS COMPOSED OF A FLAME RETARDANT MATERIAL. THE SURFACE TEMPERATURE OF THE HBT, WHEN USED IN ACCORDANCE WITH USER INSTRUCTIONS, IS DESIGNED TO BE WITHIN THE LIMITS SPECIFIED BY ISO 8185 WITH REGARD TO HOT TUBE SURFACE TEMPERATURE NOT EXCEEDING 44° CELSIUS. IT IS AN INHERENT RISK OF HBTS, THAT ADDITIONAL HEAT (ABOVE AMBIENT LEVELS) ADDED TO ANY PART OF THE TUBE VIA AN EXTERNAL SOURCE OR BEING COVERED WITH MATERIAL, MAY LEAD TO THE TUBING BECOMING DAMAGED. TO ADDRESS THIS INHERENT RISK AND AS IS REQUIRED UNDER ISO 80601, THE USER INSTRUCTIONS FOR THE 900PT562 CONTAIN THE WARNING "DO NOT ADD HEAT TO ANY PART OF THE BREATHING TUBE E.G. COVERING WITH A BLANKET, OR HEATING IT IN AN INCUBATOR OR OVERHEAD HEATER FOR A NEONATE, AS THIS COULD RESULT IN SERIOUS INJURY." ADDITIONAL "DO NOT COVER" TAGS ARE ALSO ATTACHED TO ALL HBTS TO ALERT THE USER THAT THE HBT SHOULD NOT BE COVERED. THERE ARE MANY SAFETY FEATURES INCORPORATED INTO THE AIRVO 2 SYSTEM TO PREVENT OVERHEATING AND FIRE. THESE INCLUDE: - THE HEATER WIRES IN THE HBT ARE COMPLETELY INSULATED FROM THE GAS PATH. - THE PCB AT THE PATIENT END OF THE HBT IS OVER MOULDED WITH THE THERMOPLASTIC POLYMER POLYPROPYLENE, ENSURING IT IS EXCLUDED FROM THE GAS PATH. - THE AIRVO 2 DEVICE CONTAINS TECHNOLOGY WHICH DETECTS SHORT CIRCUITS AND TRANSIENT CURRENT EVENTS IN THE HBT. WHEN DETECTED, THE AIRVO REMOVES POWER TO THE HBT. THE AIRVO 2 PERFORMS THIS DETECTION AT ANY TIME IT IS TURNED ON AND CONNECTED TO THE HBT. THIS FUNCTIONALITY IS CHECKED BY THE CONTROL SYSTEM EACH TIME THE AIRVO IS POWERED UP, OR WHEN A NEW HBT IS CONNECTED. - AN 'OVER-TEMPERATURE' SENSOR WILL AUTOMATICALLY CUT POWER TO THE MOTOR, HEATER PLATE AND HEATER WIRE IF IT DETECTS ANY OVERHEATING AT THE CHAMBER OR THE PATIENT END OF THE HBT. - THE AIRVO 2 DEVICE IS CONTINUOUSLY CHECKS POWER IN THE HBT AND DISABLES THE HEATER WIRE IF THE MEASURED POWER IS TOO HIGH.
A HEALTHCARE FACILITY IN THE UNITED KINGDOM REPORTED THAT THE HEATED BREATHING TUBE AS PART OF THE 900PT562 AIRVO TUBE AND CHAMBER KIT WITH NEBULISER ADAPTOR WAS FOUND MELTED DURING USE. THE HEALTHCARE FACILITY REPORTED THAT THE SUBJECT DEVICE HAD BEEN IN USE FOR 7 HOURS AND WAS POSITIONED ACROSS THE PATIENT'S BED DURING USE. THERE WERE NO PATIENT CONSEQUENCES.
A HEALTHCARE FACILITY IN THE UNITED KINGDOM REPORTED THAT THE HEATED BREATHING TUBE AS PART OF THE 900PT562 AIRVO TUBE AND CHAMBER KIT WITH NEBULISER ADAPTOR WAS FOUND MELTED DURING USE. THE HEALTHCARE FACILITY REPORTED THAT THE SUBJECT DEVICE HAD BEEN IN USE FOR SEVEN HOURS AND WAS POSITIONED ACROSS THE PATIENT'S BED DURING USE. THE HEALTHCARE FACILITY STATED THAT THERE WAS NO PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 459662 | FISHER & PAYKEL HEALTHCARE | HEATED BREATHING TUBE | CAF | FISHER & PAYKEL HEALTHCARE LTD | 900PT562 | NOT PROVIDED |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |