COULTER® LH 750 HEMATOLOGY ANALYZER
Report
- Report Number
- 1061932-2011-00954
- Event Type
- Malfunction
- Date Received
- July 27, 2011
- Date of Event
- June 22, 2011
- Report Date
- June 22, 2011
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- GKZ
- PMA / PMN Number
- K011342
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
BECKMAN COULTER, INC. URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IT IS NOT LIMITED TO, PROTECTIVE EYEWEAR, GLOVES, AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER.
THE CUSTOMER CONTACTED BECKMAN COULTER, INC. (BEC) TO REPORT THAT A LEAK WAS OBSERVED IN THE BOTTOM TRAY UNDER THE SECONDARY PROBE OF THEIR COULTER LH 750 HEMATOLOGY ANALYZER. THE LEAK WAS DISCOVERED WHEN THE CUSTOMER WAS PERFORMING ROUTINE CLEANING AND MAINTENANCE ON THE ANALYZER. NO PATIENT SAMPLES WERE BEING ANALYZED AT THE TIME OF THE EVENT. THERE WAS NO IMPACT TO PATIENT TREATMENT. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE - LAB COAT, GLOVES) AT THE TIME OF THE EVENT. THERE WAS NO INJURY OR EXPOSURE TO THE OPERATOR. THE OPERATOR DID NOT SEEK MEDICAL ATTENTION. THE CUSTOMER CLEANED THE LEAK AND THEN EXECUTED A STARTUP OF THE ANALYZER. THE CUSTOMER OBSERVED THAT FLUID WAS BUILDING UP IN THE BOTTOM TRAY UNDER THE SECONDARY PROBE AGAIN. THE CUSTOMER REQUESTED SERVICE TO REPAIR THE ANALYZER. THE CUSTOMER HAS AN EXPOSURE CONTROL/RISK MANAGEMENT PLAN IN PLACE. THE CUSTOMER DID NOT REVIEW THE MATERIAL SAFETY DATA SHEET (MSDS); HOWEVER, IT IS READILY AVAILABLE. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER'S SITE ON (B)(6) 2011. THE FSE REPLACED THE SECONDARY PROBE. THE ROOT CAUSE OF THIS EVENT WAS ATTRIBUTED TO A LEAK OF THE SECONDARY PROBE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COULTER® LH 750 HEMATOLOGY ANALYZER | AUTOMATED DIFFERENTIAL CELL COUNTER | GKZ | BECKMAN COULTER, INC. | NA |