30K FSI-SLI-1000 INSERT,PKD
Report
- Report Number
- 2424472-2025-00212
- Event Type
- Malfunction
- Date Received
- April 8, 2025
- Report Date
- July 11, 2025
- Manufacturer
- DENTSPLY LLC
- Product Code
- ELC
- PMA / PMN Number
- K052334/K970
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
SINCE AN OVERHEATING INSERT COULD NECESSITATE MEDICAL/SURGICAL INTERVENTION TO PRECLUDE PERMANENT DAMAGE TO A BODY STRUCTURE OR PERMANENT IMPAIRMENT OF A BODY FUNCTION, THIS MALFUNCTION WOULD BE LIKELY TO CAUSE/CONTRIBUTE TO A SERIOUS INJURY SHOULD IT RECUR. AS SUCH, THIS EVENT MEETS THE CRITERIA FOR REPORTABILITY PER 21 CFR PART 803. THE DEVICE IS AVAILABLE FOR EVALUATION, THOUGH HAS NOT BEEN RETURNED AS OF THIS REPORT. EVALUATION RESULTS WILL BE SUBMITTED AS THEY BECOME AVAILABLE.
RETURN QTY. 1, (B)(4), 30K FSI-SLI-1000-44, 00122920 BUL PER MANUFACTURE DATE. TEST METHOD / GP005-WI02 - INDUCTANCE: GAUGE ID# 5349, DUE: (B)(6) 2026, RESULT 3.11. MEETS SPEC - DOES NOT MEET SPEC, N/A. TIP CONDITION: MEETS SPEC - DOES NOT MEET SPEC, N/A. STACK CONDITION, MEETS SPEC - DOES NOT MEET SPEC, N/A. TESTED ON: G130, GAGE ID# 9626-2, DUE DATE: (B)(6) 2025, SET PRESSURE: 35 PSI. WATER FLOW: OUTPUT RATE: 35 PSI - MEETS SPEC - DOES NOT MEET SPEC, N/A. SPRAY: - MEETS SPEC - DOES NOT MEET SPEC, N/A. WATER LEAK: HP SEALING RING, PROXIMAL RING, DISTAL RING. SEAM LEAK , N/A. VIBRATION: - MEETS SPEC - DOES NOT MEET SPEC, N/A. GRIP CONDITION: MEET - DOES NOT MEET SPEC, N/A. OTHER VISUAL OBSERVATION: INSERT TIP IS BENT. INSERT WAS TESTED ON A DIGITAL THERMOMETER I.D.# 6116A. DUE DATE: (B)(6) 2025. THE TEMPERATURE WAS 82.9 °F, NO FAULT FOUND. THE SPECIFICATIONS STATE THE TEMPERATURE IS NOT TO EXCEED 118.4°F. (PER FRS-9175 REV 9). ALLEGED EVENT: CONFIRMED - NOT CONFIRMED. RETURN QTY. 1, (B)(4), 30K FSI-SLI-1000-44, 00122920 BUL PER MANUFACTURE DATE. TEST METHOD / GP005-WI02: INDUCTANCE: GAUGE ID# 5349 DUE: (B)(6) 2026 RESULT 3.05. MEETS SPEC - DOES NOT MEET SPEC - N/A. TIP CONDITION: MEETS SPEC - DOES NOT MEET SPEC, N/A. STACK CONDITION MEETS SPEC - DOES NOT MEET SPEC, N/A. TESTED ON: G130 GAGE ID# 9626-2 DUE DATE: (B)(6) 2025 SET PRESSURE: 35 PSI. WATER FLOW: OUTPUT RATE: 35 PSI MEETS SPEC - DOES NOT MEET SPEC N/A. SPRAY: MEETS SPEC - DOES NOT MEET SPEC N/A. WATER LEAK: HP SEALING RING PROXIMAL RING DISTAL RING. SEAM LEAK N/A. VIBRATION: MEETS SPEC - DOES NOT MEET SPEC N/A. GRIP CONDITION: MEET DOES NOT MEET SPEC ¿ N/A. OTHER VISUAL OBSERVATION: INSERT HAS A HP SEALING RING WATER LEAK. INSERT WAS TESTED ON A DIGITAL THERMOMETER I.D.#6116A. DUE DATE: (B)(6) 2025. THE TEMPERATURE WAS 79.6 °F, NO FAULT FOUND. THE SPECIFICATIONS STATE THE TEMPERATURE IS NOT TO EXCEED 118.4°F. (PER FRS-9175 REV 9). ALLEGED EVENT: CONFIRMED NOT CONFIRMED. RETURN QTY. 1, (B)(4), 30K FSI-SLI-1000-44, 00122920 BUL PER MANUFACTURE DATE. TEST METHOD / GP005-WI02 - INDUCTANCE: GAUGE ID# 5349 DUE: (B)(6) 2026 RESULT 3.09. MEETS SPEC - DOES NOT MEET SPEC N/A . TIP CONDITION: MEETS SPEC - DOES NOT MEET SPEC N/A. STACK CONDITION MEETS SPEC DOES NOT MEET SPEC N/A. TESTED ON: G130 GAGE ID# 9626-2 DUE DATE: (B)(6) 2025 SET PRESSURE: 35 PSI. WATER FLOW: OUTPUT RATE: 35 PSI - MEETS SPEC DOES NOT MEET SPEC N/A. SPRAY: - MEETS SPEC DOES NOT MEET SPEC N/A. WATER LEAK: HP SEALING RING PROXIMAL RING DISTAL RING. SEAM LEAK N/A. VIBRATION: MEETS SPEC DOES NOT MEET SPEC N/A. GRIP CONDITION: MEET DOES NOT MEET SPEC N/A. OTHER VISUAL OBSERVATION: INSERT IS GOOD, NO FAULT FOUND. INSERT WAS TESTED ON A DIGITAL THERMOMETER I.D.# 6116A. DUE DATE: (B)(6) 2025. THE TEMPERATURE WAS 85.4 °F, NO FAULT FOUND. THE SPECIFICATIONS STATE THE TEMPERATURE IS NOT TO EXCEED 118.4°F. (PER FRS-9175 REV 9). ALLEGED EVENT: CONFIRMED NOT CONFIRMED RETURN QTY. 1, (B)(4), 30K FSI-SLI-1000-44, 00122920 BUL PER MANUFACTURE DATE. TEST METHOD / GP005-WI02 : INDUCTANCE: GAUGE ID# 5349 DUE: (B)(6) 2026 RESULT 3.09. MEETS SPEC DOES NOT MEET SPEC N/A. TIP CONDITION: MEETS SPEC DOES NOT MEET SPEC N/A. STACK CONDITION MEETS SPEC DOES NOT MEET SPEC N/A TESTED ON: G130 GAGE ID# 9626-2 DUE DATE: (B)(6) 2025 SET PRESSURE: 35 PSI. WATER FLOW: OUTPUT RATE: 35 PSI - MEETS SPEC DOES NOT MEET SPEC N/A. SPRAY: - MEETS SPEC DOES NOT MEET SPEC N/A. WATER LEAK: HP SEALING RING PROXIMAL RING DISTAL RING. SEAM LEAK N/A. VIBRATION: MEETS SPEC DOES NOT MEET SPEC N/A. GRIP CONDITION: MEET DOES NOT MEET SPEC N/A. OTHER VISUAL OBSERVATION: INSERT IS GOOD, NO FAULT FOUND. INSERT WAS TESTED ON A DIGITAL THERMOMETER I.D.#6116A. DUE DATE: (B)(6) 2025. THE TEMPERATURE WAS 85.4 °F, NO FAULT FOUND. THE SPECIFICATIONS STATE THE TEMPERATURE IS NOT TO EXCEED 118.4°F. (PER FRS-9175 REV 9). ALLEGED EVENT: CONFIRMED NOT CONFIRMED. RETURN QTY. 1, (B)(4), 30K FSI-SLI-1000-44, 00122920 BUL PER MANUFACTURE DATE. TEST METHOD / GP005-WI02 : INDUCTANCE: GAUGE ID# 5349 DUE: (B)(6) 2026 RESULT 3.09. MEETS SPEC DOES NOT MEET SPEC N/A. TIP CONDITION: MEETS SPEC DOES NOT MEET SPEC N/A. STACK CONDITION MEETS SPEC DOES NOT MEET SPEC N/A. TESTED ON: G130 GAGE ID# 9626-2 DUE DATE: (B)(6) 2025 SET PRESSURE: 35 PSI. WATER FLOW: OUTPUT RATE: 35 PSI MEETS SPEC DOES NOT MEET SPEC N/A. SPRAY: MEETS SPECDOES NOT MEET SPEC N/A. WATER LEAK: HP SEALING RING PROXIMAL RING DISTAL RING. SEAM LEAK N/A. VIBRATION: MEETS SPEC DOES NOT MEET SPEC N/A. GRIP CONDITION: MEET DOES NOT MEET SPEC N/A. OTHER VISUAL OBSERVATION: INSERT IS GOOD, NO FAULT FOUND. INSERT WAS TESTED ON A DIGITAL THERMOMETER I.D.# 6116A. DUE DATE: (B)(6) 2025. THE TEMPERATURE WAS 85.4 °F, NO FAULT FOUND. THE SPECIFICATIONS STATE THE TEMPERATURE IS NOT TO EXCEED 118.4°F. (PER FRS-9175 REV 9). ALLEGED EVENT: CONFIRMED NOT CONFIRMED. RETURN QTY. 1, (B)(4), 30K FSI-SLI-1000-44, 00122920 BUL PER MANUFACTURE DATE. TEST METHOD / GP005-WI02: INDUCTANCE: GAUGE ID# 5349 DUE: (B)(6) 2026 RESULT 3.09. MEETS SPEC DOES NOT MEET SPEC N/A. TIP CONDITION: MEETS SPEC DOES NOT MEET SPEC N/A. STACK CONDITION MEETS SPEC DOES NOT MEET SPEC N/A. TESTED ON: G130 GAGE ID# 9626-2 DUE DATE: (B)(6) 2025 SET PRESSURE: 35 PSI. WATER FLOW: OUTPUT RATE: 35 PSI MEETS SPEC DOES NOT MEET SPEC N/A. SPRAY: MEETS SPEC DOES NOT MEET SPEC N/A. WATER LEAK: HP SEALING RING PROXIMAL RING DISTAL RING. SEAM LEAK N/A. VIBRATION: MEETS SPEC DOES NOT MEET SPEC N/A. GRIP CONDITION: MEET DOES NOT MEET SPEC N/A. OTHER VISUAL OBSERVATION: INSERT HAS A HP SEALING RING WATER LEAK, APPEARS TO BE MISSING THE SEALING RING. ALLEGED EVENT: CONFIRMED NOT CONFIRMED.
IN THIS EVENT IT IS REPORTED THAT A 30K FSI-SLI-1000 INSERT,PKD ALLEGEDLY GET HOT AND THERE IS A WATER ISSUE. NO INJURY OCCURRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1238619 | 30K FSI-SLI-1000 INSERT,PKD | SCALER, ULTRASONIC | ELC | DENTSPLY LLC | 00122920 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |