FDA Adverse Event Malfunction Summary report: N

ENDOWRIST SP

MDR report key: 21800907 · Received April 8, 2025

Report

Report Number
2955842-2025-13935
Event Type
Malfunction
Date Received
April 8, 2025
Date of Event
March 4, 2024
Report Date
March 13, 2025
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
PMA / PMN Number
K182371
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL INC. (ISI) RECEIVED THE SINGLE-PORT (SP) ROUND TOOTH RETRACTOR (RTR) INSTRUMENT FOR FAILURE ANALYSIS INVESTIGATION. THE INSTRUMENT WAS ANALYZED, AND THE COMPLAINT COULD NOT BE REPLICATED NOR CONFIRMED. THE INSTRUMENT WAS TESTED ON AN IN-HOUSE SYSTEM. THE INSTRUMENT PASSED THE RECOGNITION AND ENGAGEMENT TESTS. THE INSTRUMENT MOVED INTUITIVELY WITH FULL RANGE OF MOTION IN ALL DIRECTIONS. THE GRIP TIPS OPENED AND CLOSED PROPERLY. THE PROBABLE ROOT CAUSE CANNOT BE DETERMINED BASED ON THE INFORMATION PROVIDED AND FAILURE ANALYSIS RESULTS.

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) HAS ISSUED A RETURN MATERIAL AUTHORIZATION (RMA) TO HAVE THE DEVICE RETURNED. HOWEVER, ISI HAS NOT RECEIVED THE INSTRUMENT INVOLVED WITH THE ALLEGED ISSUE TO PERFORM FAILURE ANALYSIS.

Description of Event or Problem · 0

REFER TO H11 FOR FOLLOW-UP INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SIGMOID COLECTOMY SURGICAL PROCEDURE, SINGLE-PORT (SP) ROUND TOOTH RETRACTOR (RTR) INSTRUMENT MOVED NON-INTUITIVELY. NO FRAGMENT FELL INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
740957 ENDOWRIST SP ROUND TOOTH RETRACTOR NAY INTUITIVE SURGICAL, INC 430002-57 U10220512 0020

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES