ENDOWRIST SP
Report
- Report Number
- 2955842-2025-13935
- Event Type
- Malfunction
- Date Received
- April 8, 2025
- Date of Event
- March 4, 2024
- Report Date
- March 13, 2025
- Manufacturer
- INTUITIVE SURGICAL, INC
- Product Code
- NAY
- PMA / PMN Number
- K182371
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TW
- Reporter Occupation
- 003
Narratives
INTUITIVE SURGICAL INC. (ISI) RECEIVED THE SINGLE-PORT (SP) ROUND TOOTH RETRACTOR (RTR) INSTRUMENT FOR FAILURE ANALYSIS INVESTIGATION. THE INSTRUMENT WAS ANALYZED, AND THE COMPLAINT COULD NOT BE REPLICATED NOR CONFIRMED. THE INSTRUMENT WAS TESTED ON AN IN-HOUSE SYSTEM. THE INSTRUMENT PASSED THE RECOGNITION AND ENGAGEMENT TESTS. THE INSTRUMENT MOVED INTUITIVELY WITH FULL RANGE OF MOTION IN ALL DIRECTIONS. THE GRIP TIPS OPENED AND CLOSED PROPERLY. THE PROBABLE ROOT CAUSE CANNOT BE DETERMINED BASED ON THE INFORMATION PROVIDED AND FAILURE ANALYSIS RESULTS.
INTUITIVE SURGICAL, INC. (ISI) HAS ISSUED A RETURN MATERIAL AUTHORIZATION (RMA) TO HAVE THE DEVICE RETURNED. HOWEVER, ISI HAS NOT RECEIVED THE INSTRUMENT INVOLVED WITH THE ALLEGED ISSUE TO PERFORM FAILURE ANALYSIS.
REFER TO H11 FOR FOLLOW-UP INFORMATION.
IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SIGMOID COLECTOMY SURGICAL PROCEDURE, SINGLE-PORT (SP) ROUND TOOTH RETRACTOR (RTR) INSTRUMENT MOVED NON-INTUITIVELY. NO FRAGMENT FELL INSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 740957 | ENDOWRIST SP | ROUND TOOTH RETRACTOR | NAY | INTUITIVE SURGICAL, INC | 430002-57 | U10220512 0020 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | DA VINCI INSTRUMENTS AND ACCESSORIES |