FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2180036 · Received July 27, 2011

Report

Report Number
2531779-2011-05269
Event Type
Malfunction
Date Received
July 27, 2011
Report Date
June 28, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THE FOLLOWING WAS INVESTIGATED ONCE THE PUMP WAS RETURNED: TESTING CONFIRMED INTERMITTENT RESPONSES TO BUTTON PRESSES ON ALL THE BUTTONS ON THE KEYPAD. CONFIRMED BUTTONS ARE STICKING AND PRODUCING SCROLLING ACTION. REMOVED KEYPAD COVER TO CHECK CONDITION OF THE BUTTON CONTACTS. CONTAMINATION WAS PRESENT UNDER THE CONTACTS OF ALL BUTTONS. OBSERVED ALL THE BUTTON CONTACTS ARE INVERTED.

Description of Event or Problem · 1

THE PATIENT CLAIMED THAT ALL BUTTONS REQUIRED SEVERAL PRESSES TO ACTIVATE THE DESIRED PUMP FUNCTIONS. THE KEYPAD IS REPORTEDLY INTACT AND HAS NOT BEEN EXPOSED TO MOISTURE. THERE WAS NO ADVERSE EVENT ASSOCIATE WITH THIS COMPLAINT. THE PUMP WAS REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 61 YR