FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2180002 · Received July 27, 2011

Report

Report Number
2531779-2011-05266
Event Type
Malfunction
Date Received
July 27, 2011
Report Date
June 28, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS AND EVALUATED. THE PUMP WAS INTERMITTENTLY UNRESPONSIVE TO UP AND DOWN KEYPAD BUTTON PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE WAS FOUND UNDER THE BUTTON CONTACTS.

Description of Event or Problem · 1

ON (B)(6) 2011, THE PATIENT CONTACTED ANIMAS ALLEGING THAT KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE PATIENT DID NOT ALLEGE ANY HARM OR INJURY BECAUSE OF THE REPORTED ISSUE. BASED ON THE INFORMATION PROVIDED, THIS COMPLAINT IS BEING REPORTED DUE TO THE FOLLOWING CONCLUSION: THE PATIENT CLAIMED THAT THE PUMP WAS UNRESPONSIVE TO UP AND DOWN BUTTON PRESSES. THERE IS NO EVIDENCE HOWEVER THAT THE ALLEGED ISSUE CONTRIBUTED TO A SERIOUS INJURY. THE PATIENT DID NOT REPORT ANY BLOOD GLUCOSE VALUES, SYMPTOMS, OR TREATMENT SUGGESTIVE OF A SERIOUS INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 19 YR