FDA Adverse Event Injury Summary report: N

ENDURE

MDR report key: 21796641 · Received April 8, 2025

Report

Report Number
21796641
Event Type
Injury
Date Received
April 8, 2025
Date of Event
January 25, 2025
Report Date
April 2, 2025
Manufacturer
ENDURE INDUSTRIES, INC
Product Code
BTQ
Product Problem
Yes
Report Source
User Facility report
Reporter Location
MD
Reporter Occupation
RISK MANAGER
Health Professional
N

Narratives

Description of Event or Problem · 0

DELAY IN RECOGNITION OF RETAINED NASOPHARYNGEAL AIRWAY (NPA) PLACED PRIOR TO ADMISSION CAUSING BLEEDING, SWELLING, NEED FOR EMERGENT TRACH WHEN PATIENT WAS EXTUBATED AND REQUIRED REINTUBATION. NPA WAS CLEAR AND HAS NO RADIOPAQUE STRIP MAKING IT MORE DIFFICULT TO RECOGNIZE BOTH VISUALLY AND RADIOGRAPHICALLY. UNABLE TO IDENTIFY MANUFACTURER OR ANY DEVICE INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
754406 ENDURE AIRWAY, NASOPHARYNGEAL BTQ ENDURE INDUSTRIES, INC NPS3021217

Patients

Seq Age Sex Outcome Treatment
1 67 YR Male Hospitalization| R