FDA Adverse Event Malfunction Summary report: N

TPRLC 133 T1 PPS HO 17X154MM

MDR report key: 21796065 · Received April 8, 2025

Report

Report Number
0001825034-2025-01018
Event Type
Malfunction
Date Received
April 8, 2025
Date of Event
March 14, 2025
Report Date
July 1, 2025
Manufacturer
ZIMMER BIOMET, INC.
Product Code
LZO
PMA / PMN Number
K200196
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED: B4;D9;G3;H2;H3;H6 PRODUCT WAS RETURNED AND EVALUATED. VISUAL EVALUATION OF THE RETURNED PRODUCT/PROVIDED PHOTOS IDENTIFIED DAMAGE TO THE STERILE PACKAGING (BLISTER). STERILITY HAD NOT BEEN COMPROMISED. REVIEW OF THE DEVICE HISTORY RECORDS IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING RELATED TO THE REPORTED EVENT. REPORTED EVENT IS NOT RELATED TO A COMBINATION OF PRODUCTS; THEREFORE, A COMPATIBILITY REVIEW IS NOT APPLICABLE. THE REPORTED EVENT DID NOT OCCUR IN AN OPERATING ROOM OR AS PART OF A MEDICAL PROCEDURE; MEDICAL RECORDS ARE NOT AVAILABLE FOR REVIEW. THE CONDITION OF THE DEVICE WHEN IT LEFT ZIMMER BIOMET IS CONSIDERED CONFORMING TO SPECIFICATION. THE ROOT CAUSE OF THE REPORTED EVENT CAN BE ATTRIBUTED TO TRANSIT DAMAGE AND A PACKAGING DESIGN ISSUE. THE REPORTED EVENT HAS BEEN CONFIRMED BY EVALUATION OF THE RETURNED PRODUCT AND PROVIDED PHOTOS. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.

Additional Manufacturer Narrative · 0

(B)(4). D10: 51-104140 ITEM NAME TPRLC 133 T1 PPS HO 14X148MM LOT # 6551110 51-145170 ITEM NAME TPRLC XR MP T1 PPS 17X119MM LOT # 6639629 51-104140 ITEM NAME TPRLC 133 T1 PPS HO 14X148MM LOT # 7008536 51-107170 ITEM NAME TPRLC 133 MP TYPE1 PPS HO 17.0 LOT # 6700680 51-104170 ITEM NAME TPRLC 133 T1 PPS HO 17X154MM LOT # 6820307 51-103160 ITEM NAME TPRLC 133 T1 PPS SO 16X152MM LOT # 6143297 G2: FOREIGN: JAPAN CUSTOMER HAS INDICATED THAT THE PRODUCT IS IN PROCESS OF BEING RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 0

NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.

Description of Event or Problem · 0

IT WAS REPORTED THAT WHILE INSPECTING THE PRODUCT, THE STERILE PACKAGING WAS FOUND DAMAGED. THERE WAS NO PATIENT INVOLVEMENT. THERE IS NO FURTHER INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
463918 TPRLC 133 T1 PPS HO 17X154MM PROSTHESIS, HIP LZO ZIMMER BIOMET, INC. 6985695

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown PLEASE SEE H11.