FDA Adverse Event
Malfunction
Summary report: N
NC 1 CM/WITH OPMI: CS NC
MDR report key: 217939
·
Received April 1, 1999
Report
- Report Number
- 2431026-1999-00001
- Event Type
- Malfunction
- Date Received
- April 1, 1999
- Date of Event
- February 22, 1999
- Report Date
- March 24, 1999
- Manufacturer
- CARL ZEISS, INC.
- Product Code
- FSY
- Removal / Correction Number
- NOT RECEIVED YET
- Report Source
- Manufacturer report
- Reporter Location
- LA, US
- Reporter Occupation
- NURSE
Narratives
Description of Event or Problem · 1
DURING THE BEGINNING OF A PROCEDURE IN THE OPERATING ROOM, THE STAFF MOVED THE NC 1 CEILING MOUNT INTO THE FIELD. AFTER MOVING IT A LIGHT WAS NOT OPERATIONAL; THEY MOVED IT AWAY TO CHECK IT AND THE UNIT FELL. NO ONE WAS DIRECTLY HIT, BUT A NURSE ALLEGEDLY SUFFERED A MINOR INJURY AS THE UNIT ROLLED ON THE FLOOR. THIS UNIT WAS A CONVERTED FLOOR STAND THAT REQUIRED A PIN TO STOP THE CLUTCH/SCREW FROM BACKING OUT. THE PIN WAS NOT IN PLACE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NC 1 CM/WITH OPMI: CS NC | CEILING MOUNT SUSPENSION SYSTEM | FSY | CARL ZEISS, INC. | * | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |