FDA Adverse Event Malfunction Summary report: N

R SERIES

MDR report key: 21786049 · Received April 7, 2025

Report

Report Number
1220908-2025-01163
Event Type
Malfunction
Date Received
April 7, 2025
Report Date
March 17, 2025
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
UDI-DI
00847946019693
PMA / PMN Number
P160022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS EVALUATED BY ZOLL MEDICAL CORPORATION. THE CUSTOMER'S REPORT WAS DUPLICATED AND ATTRIBUTED TO THE MISALIGNED FLEX CABLE TO A CONNECTOR. THE FLEX CABLE WAS RESEATED TO RESOLVE THE REPORT. THE DEVICE WAS RECERTIFIED AND RETURNED TO THE CUSTOMER. ANALYSIS OF REPORTS OF THIS TYPE HAS NOT IDENTIFIED AN INCREASE IN TREND.

Additional Manufacturer Narrative · 0

ZOLL MEDICAL CORPORATION HAS RECEIVED THE PRODUCT AND WILL BE PROVIDING A SUPPLEMENTAL REPORT WHEN OUR INVESTIGATION IS COMPLETED.

Description of Event or Problem · 0

COMPLAINANT ALLEGED THAT DURING BIOMED TESTING, THE DEVICE POWERED UP IN INCORRECT MODE. COMPLAINANT INDICATED THAT THERE WAS NO PATIENT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1458376 R SERIES DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORPORATION 30001009001310012 NA 00847946019693

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown