CARDIOSAVE HYBRID, TYPE B PLUG
Report
- Report Number
- 2249723-2025-0001622
- Event Type
- Malfunction
- Date Received
- April 7, 2025
- Date of Event
- March 24, 2025
- Report Date
- December 16, 2025
- Manufacturer
- DATASCOPE CORP.
- Product Code
- DSP
- PMA / PMN Number
- K112372
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
- Health Professional
- Yes
Narratives
A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF OUR INVESTIGATION.
UPDATED FIELDS: B4,D8, D9, E3, G1(CONTACT PERSON ¿ MFG SITE), G3, G6, H2,H3, H6(INVESTIGATION FINDINGS, TYPE OF INVESTIGATION, INVESTIGATION CONCLUSIONS,COMPONENT CODE), H1. THE FAILURE ANALYSIS AND TESTING DEPT. RECEIVED PART NUMBER 0997-00-0565 WITH A REPORTED UNIT FAILURE OF A HELIUM LEAK. THE FAT PERFORMED A VISUAL INSPECTION AND FOUND THE PART TO BE OXIDIZED IN SOME AREAS. SEE ATTACHMENT. THE FAT INSTALLED THE PART IN CARDIOSAVE TEST FIXTURE SERIAL NUMBER (B)(6) AND TESTED THE PART TO FACTORY SPECIFICATIONS PER THE CARDIOSAVE SERVICE MANUAL PART NUMBER 0070-00-0639 REVISION R. CONFIRMED THE REPORTED HELIUM LEAK. NO ROOT CAUSE CAN BE IDENTIFIED. PROBABLE IS THE OXIDATION ON THE PART. RETAINING THE PART IN THE FAILURE ANALYSIS AND TESTING DEPARTMENT PER PROCEDURE NUMBER (B)(6) REV. AU. A GETINGE FIELD SERVICE ENGINEER (FSE) EVALUATED THE UNIT FOR THE REPORTED MALFUNCTION. THE FSE REPLACED THE HELIUM RESERVOIR ASSEMBLY (0997-00-0565). THE UNIT PASSED ALL FUNCTIONAL AND SAFETY TESTS PER MANUFACTURER SPECIFICATIONS.
UPDATED FIELD : B4 , G3 , G6 , H2 , H11. CORRECTED FIELD: H6 ( INVESTIGATION FINDINGS).
UPDATED FIELDS: B4, G3, G6, H2, H11. CORRECTED FIELDS: B5, H6 (INVESTIGATION FINDINGS).
IT WAS REPORTED THAT THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) HAD EXCESSIVE LEAK OF THE HELIUM. NO PATIENT INVOLVEMENT WAS REPORTED.
IT WAS REPORTED THAT PRIOR TO USE, THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) HAD EXCESSIVE LEAK OF THE HELIUM. THERE WAS NO PATIENT INVOLVEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1569951 | CARDIOSAVE HYBRID, TYPE B PLUG | SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL | DSP | DATASCOPE CORP. | 0998-00-0800-53 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |