FDA Adverse Event Malfunction Summary report: N

CARDIOSAVE HYBRID, TYPE B PLUG

MDR report key: 21783926 · Received April 7, 2025

Report

Report Number
2249723-2025-0001622
Event Type
Malfunction
Date Received
April 7, 2025
Date of Event
March 24, 2025
Report Date
December 16, 2025
Manufacturer
DATASCOPE CORP.
Product Code
DSP
PMA / PMN Number
K112372
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A SUPPLEMENTAL REPORT WILL BE SUBMITTED UPON COMPLETION OF OUR INVESTIGATION.

Additional Manufacturer Narrative · 0

UPDATED FIELDS: B4,D8, D9, E3, G1(CONTACT PERSON ¿ MFG SITE), G3, G6, H2,H3, H6(INVESTIGATION FINDINGS, TYPE OF INVESTIGATION, INVESTIGATION CONCLUSIONS,COMPONENT CODE), H1. THE FAILURE ANALYSIS AND TESTING DEPT. RECEIVED PART NUMBER 0997-00-0565 WITH A REPORTED UNIT FAILURE OF A HELIUM LEAK. THE FAT PERFORMED A VISUAL INSPECTION AND FOUND THE PART TO BE OXIDIZED IN SOME AREAS. SEE ATTACHMENT. THE FAT INSTALLED THE PART IN CARDIOSAVE TEST FIXTURE SERIAL NUMBER (B)(6) AND TESTED THE PART TO FACTORY SPECIFICATIONS PER THE CARDIOSAVE SERVICE MANUAL PART NUMBER 0070-00-0639 REVISION R. CONFIRMED THE REPORTED HELIUM LEAK. NO ROOT CAUSE CAN BE IDENTIFIED. PROBABLE IS THE OXIDATION ON THE PART. RETAINING THE PART IN THE FAILURE ANALYSIS AND TESTING DEPARTMENT PER PROCEDURE NUMBER (B)(6) REV. AU. A GETINGE FIELD SERVICE ENGINEER (FSE) EVALUATED THE UNIT FOR THE REPORTED MALFUNCTION. THE FSE REPLACED THE HELIUM RESERVOIR ASSEMBLY (0997-00-0565). THE UNIT PASSED ALL FUNCTIONAL AND SAFETY TESTS PER MANUFACTURER SPECIFICATIONS.

Additional Manufacturer Narrative · 0

UPDATED FIELD : B4 , G3 , G6 , H2 , H11. CORRECTED FIELD: H6 ( INVESTIGATION FINDINGS).

Additional Manufacturer Narrative · 0

UPDATED FIELDS: B4, G3, G6, H2, H11. CORRECTED FIELDS: B5, H6 (INVESTIGATION FINDINGS).

Description of Event or Problem · 0

IT WAS REPORTED THAT THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) HAD EXCESSIVE LEAK OF THE HELIUM. NO PATIENT INVOLVEMENT WAS REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO USE, THE CARDIOSAVE INTRA-AORTIC BALLOON PUMP (IABP) HAD EXCESSIVE LEAK OF THE HELIUM. THERE WAS NO PATIENT INVOLVEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1569951 CARDIOSAVE HYBRID, TYPE B PLUG SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL DSP DATASCOPE CORP. 0998-00-0800-53

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown