FDA Adverse Event Injury Summary report: N

STYLE 15 SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 21782662 · Received April 6, 2025

Report

Report Number
9617229-2025-05595
Event Type
Injury
Date Received
April 6, 2025
Date of Event
January 29, 2010
Report Date
June 1, 2025
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: D9, H3, H6. DEVICE EVALUATION: BASED ON THE PRODUCT ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS ARE: RUPTURE: OBSERVED, OPENING ASSESSED AS FOLD FLAW OPENING. PAIN: UNABLE TO OBSERVE AS IT IS NOT RELATED TO THE MANUFACTURING PROCESS. AS PER THE INVESTIGATION PROCEDURE, CREASE, WEAR ABRASION AND NON-PENETRATING NICK WERE COMPLETED AND NONE OF THE OBSERVATIONS ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, NO FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 0

THIS IS A FOLLOW-UP REPORT TO A MEDWATCH SUBMITTED UNDER MANUFACTURE REPORT NUMBER 9617229-2024-0011441. A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN INITIATED. IF ANY NEW, CHANGED OR CORRECTED INFORMATION IS NOTED, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.

Description of Event or Problem · 0

PATIENT REPORTED "PAIN". HEALTHCARE PROFESSIONAL CONFIRMED RIGHT SIDE INTRACAPSULAR RUPTURE DIAGNOSED VIA MRI. DEVICE HAS BEEN EXPLANTED AND REPLACED.

Description of Event or Problem · 0

PATIENT REPORTED "PAIN". HEALTHCARE PROFESSIONAL CONFIRMED RIGHT SIDE INTRACAPSULAR RUPTURE DIAGNOSED VIA MRI. DEVICE HAS BEEN EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1736383 STYLE 15 SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 1773851

Patients

Seq Age Sex Outcome Treatment
1 47 YR Female Required Intervention