FDA Adverse Event Injury Summary report: N

INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 21778549 · Received April 4, 2025

Report

Report Number
9617229-2025-05550
Event Type
Injury
Date Received
April 4, 2025
Date of Event
March 4, 2025
Report Date
July 13, 2025
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT OF CAPSULAR CONTRACTURE, VISIBILITY/PALPABILITY AND BREAST PAIN WAS RECEIVED ON MAY 29, 2025, WITH LOT NUMBER 1230960. BASED ON THE PRODUCT ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS ARE: CAPSULAR CONTRACTURE: UNABLE TO OBSERVE AS IT IS NOT RELATED TO THE DEVICE. VISIBILITY/PALPABILITY: UNABLE TO OBSERVE AS IT IS NOT RELATED TO THE DEVICE. BREAST PAIN: UNABLE TO OBSERVE AS IT IS NOT RELATED TO THE DEVICE. AS PER THE INVESTIGATION PROCEDURES, CREASES AND WEAR ABRASION WERE COMPLETED AND NONE OF THE OBSERVATIONS ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, NO FURTHER ACTIONS ARE REQUIRED. PEER/ SUPERVISOR REVIEWER.

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. THE EVENT OF "CAPSULAR CONTRACTURE, BAKER GRADE III" IS A PHYSIOLOGICAL COMPLICATION AND ANALYSIS OF THE DEVICE GENERALLY DOES NOT ASSIST ALLERGAN IN DETERMINING A PROBABLE CAUSE FOR THIS EVENT. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: CAPSULAR CONTRACTURE, BAKER GRADE III.

Additional Manufacturer Narrative · 0

ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: B5, D6B, H6.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED "CAPSULAR CONTRACTURE, BAKER GRADE III AND BREAST FEEL TIGHT AND HURTS". THE DEVICE HAS BEEN EXPLANTED AND REPLACED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED "CAPSULAR CONTRACTURE, BAKER GRADE III AND BREAST FEEL TIGHT AND HURTS". THIS RECORD IS FOR THE RIGHT SIDE. DEVICE REMAINS IMPLANTED.

Description of Event or Problem · 0

THE DEVICE HAS BEEN EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
647120 INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 1230960

Patients

Seq Age Sex Outcome Treatment
1 36 YR Female Required Intervention