M-SERIES (DC) HEATED HUMIDIFIER
Report
- Report Number
- 2518422-2011-00075
- Event Type
- Malfunction
- Date Received
- July 6, 2011
- Date of Event
- June 1, 2011
- Report Date
- June 7, 2011
- Manufacturer
- RESPIRONICS, INC.
- Product Code
- BZD
- PMA / PMN Number
- K072996
- Removal / Correction Number
- Z-1260-2009
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
THE MFR COMPLETED THE INVESTIGATION OF THE FAILURE MODE OF THE J3 CONNECTOR AND CONCLUDED IT IS CAUSED BY HIGH RESISTANCE HEATING BETWEEN THE HEATER-PLATE CONNECTOR AND THE HEATER-PLATE WIRE HARNESS. ENGINEERING DETERMINED THAT THE HIGH RESISTANCE HEATING WAS ATTRIBUTABLE TO CORROSION OF THE CONNECTOR'S CONTACT TERMINALS DUE TO FRETTING AND/OR INTERMITTENT ELECTRICAL CONNECTION CAUSED BY INSUFFICIENT SPRING FORCES ON THE WIRE-HARNESS. (CONTINUED ON MFR # 251842-2011-00061, P. 3 OF 3). METHODS, HISTORICAL REVIEW OF AFFECTED COMPLAINT RECORDS. FRETTING IS A COMBINATION OF FRICTIONAL WEAR AND CORROSION CAUSED BY SMALL AMPLITUDE MOTION BETWEEN THE MATING MATERIALS OF THE CONNECTOR. THE SOURCE OF THE SMALL AMPLITUDE MOTION WAS ATTRIBUTED TO TEMPERATURE CYCLING BETWEEN THE MATING METALS DUE TO ELECTRICAL CURRENT, PULSED TO THE HEATER LOAD. CORROSION GENERATED ON THE TERMINAL CONTACTS DUE TO FRETTING RESULTED IN INCREASED RESISTANCE WITHIN THE CONNECTOR CAUSING INCREASED HEAT. THE INSUFFICIENT SPRING FORCE ESTABLISHED BY THE FEMALE TERMINAL CONTACTS WITHIN THE HEATER PLATE WIRE-HARNESS IS ATTRIBUTABLE TO THE GEOMETRY OF THE TERMINAL CONTACTS DEVELOPED DURING THE SUPPLIER'S MANUFACTURING PROCESS. THE INTERMITTENT CONNECTION CAN POTENTIALLY CAUSE AN INCREASE N HEAT DUE TO ELECTRICAL ARCING. THE HEATER PLATE WIRE-HARNESS CONNECTOR HAD BEEN PURCHASED FROM TWO SEPARATE SUPPLIERS. BASED ON AN ANALYSIS OF THE MFR'S COMPLAINT DATABASE AND EVAL OF THE UNITS RETURNED, THE MFR DETERMINED THAT THE FAILURE MODE WAS ISOLATED TO A SINGLE SUPPLIER. THE COMPOSITION OF THE METAL TERMINAL CONTACTS WITHIN ONE SUPPLIER'S CONNECTOR PROVED TO BE SUSCEPTIBLE TO FRETTING AND THE GEOMETRY OF THE CONDUCTORS FROM THE SAME SUPPLIER'S CONNECTOR INCREASED THE POTENTIAL FOR AN INTERMITTENT CONNECTION TO OCCUR. CONNECTORS PURCHASED FROM THE OTHER SUPPLIER DID NOT HAVE THE SAME CHARACTERISTICS IN GEOMETRY OR THE COMPOSITION OF METAL TERMINAL CONTACTS. THE MFR WAS ABLE TO ISOLATE THE POPULATION OF DEVICES POTENTIALLY MANUFACTURED WITH THE SPECIFIC SUPPLIER'S SUSPECT CONNECTOR AND ISSUED A VOLUNTARY FIELD CORRECTIVE ACTION. THE FDA WAS NOTIFIED IN JULY 2009 AND PHILIPS RESPIRONICS WAS ISSUED RECALL NUMBER Z-1260-2009.
A DURABLE MEDICAL EQUIPMENT SUPPLIER (DME) REPORTED THAT A PHILIPS RESPIRONICS HEATED HUMIDIFIER AND AN ASSOCIATED CONTINUOUS POSITIVE AIRWAY PRESSURE (CPAP) DEVICE HAD MELTED TOGETHER. THE PT ALLEGED THAT THE DEVICE HAD CAUGHT ON FIRE BUT REPORTED THAT THERE WERE NO FLAMES OBSERVED. THERE WAS NO REPORT OF INJURY OR HARM. THE DEVICE WAS EXAMINED BY THE MFR AND THE COMPLAINT THE HEATED HUMIDIFIER HAD EMITTED SMOKE AND MELTED TO IS ASSOCIATED BASE UNIT WAS CONFIRMED. A THERMAL VOID WAS FOUND IN BOTH THE DEVICE'S TOP AND BOTTOM ENCLOSURES. THE DAMAGE TO THE ENCLOSURES WAS PROXIMAL TO A THERMALLY DAMAGED PORTION OF THE DEVICE'S PRINTED CIRCUIT ASSEMBLY (PCA), IN THE LOCATION OF THE J3 CONNECTOR TO THE HUMIDIFIER HEATER-PLATE. THE ENCLOSURE OF THE ASSOCIATED CPAP WAS NOT COMPROMISED AND CPAP FUNCTION WAS UNAFFECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | M-SERIES (DC) HEATED HUMIDIFIER | CPAP SYSTEM | BZD | RESPIRONICS, INC. | 1051158 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR | REMSTAR M-SERIES: 100M, (B)(4) |