SZ 32 MM PINNACLE ALTRX POLYETHYLENE ACETABULAR
Report
- Report Number
- 2176901
- Event Type
- Injury
- Date Received
- July 19, 2011
- Date of Event
- May 3, 2011
- Report Date
- July 14, 2011
- Manufacturer
- JNJ DEPUY ORTHOPEDICS
- Product Code
- LPH
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- PA, US
- Reporter Occupation
- RISK MANAGER
Narratives
PT WITH HISTORY OF DEGENERATIVE ARTHRITIS REQUIRING A RIGHT HIP REPLACEMENT. INITIAL SURGERY ON (B)(6) 2011 WAS A RIGHT TOTAL HIP REPLACEMENT USING AN S-ROM SYSTEM WITH A 15 X 20 STEM LONG NECK OFFSET WITH A 32 BEARING HEAD, COBALT CHROME +3 AND A 54 CUP WITH A POLYETHYLENE 32 MM BEARING SURFACE PUT IN THE 9 O'CLOCK POSITION. INITIAL POST OPERATIVE X-RAY, AND FOLLOW UP X-RAYS SHOW GOOD POSITION. ON (B)(6) 2011, PT COMPLAINS OF PAIN IN GROIN. PT STATES THAT HE NOTICED A "NOISE" IN THE JOINT AND FELT A POP. HE COULDN'T PUT MUCH WEIGHT ON IT. X-RAY TAKEN WAS SHOWING THAT THE POLYETHYLENE IS EITHER SPLIT AND WORN OR SPLIT OR DISLODGED ITSELF FROM ITS FIXATION AND SPUN. ON (B)(6) 2011, PT WAS RETURNED TO OPERATING ROOM FOR REVISION OF POLYETHYLENE AND FEMORAL HEAD TO RIGHT TOTAL HIP REPLACEMENT. FINDING DURING THE PROCEDURE SHOWED DISASSOCIATION POLYETHYLENE LINER, WITH EVIDENCE THAT THE LOCKING TINES HAD FAILED IN 3 LOCATIONS, AND THE HEAD WAS SCRATCHED ON THE BEARING SURFACE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SZ 32 MM PINNACLE ALTRX POLYETHYLENE ACETABULAR | HIP IMPLANT (IMPLANT ON THE FLY) | LPH | JNJ DEPUY ORTHOPEDICS | 1221-32-254 | FB3A71 | |
| 2 | SLSZ 32 MM PINNACLE ALTRX POLYETHYLENE ACETABULAR | HIP IMPLANT (IMPLANT ON THE FLY) | KWA | JNJ DEPUY ORTHOPEDICS | FC1FG1 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR | Required Intervention | FEMORAL STEM, MANUFACTURER: JNJ DEPUY ORTHOPEDICS| IMPLANT NAME: SZ +3 S-ROM, MANUFACTURER:| INVENTORY ITEM: IMPLANT ON THE FLY ,| IMPLANT NAME: SZ 20 F XXL S-ROM PROXIMAL SLEEVE| SZ 20X15 X 165, 36 STD NECK +8 LATERAL S-ROM| SCREW CANC PINNACLE 6.5X25MM,| MODEL/CAT NO.: 121725500| SCREW CANC PINNACLE 6.5X20MM - (B)(4)| JNJ DEPUY ORTH, LOT NO: D00447427| MODEL/CAT NO.: 121720500| LOG250614, MANUFACTURER: JNJ DEPUY ORTHOPEDICS,| INVENTORY ITEM: IMPLANT ON THE FLY ,| SERIAL NO.: MODEL/CAT NO.: (B)(4),| INVENTORY ITEM: IMPLANT ON THE FLY,| INVENTORY ITEM: IMPLANT ON THE FLY ,| LOT NO.: 3193894| SERIAL NO.: MODEL/CAT NO.: (B)(4), IMPLANT NAME:| SERIAL NO.: MODEL/CAT NO.: (B)(4),| SCREW CANC PINNACLE 6.5X20MM,| SERIAL NO.: MODEL/CAT NO.: (B)(4),| LOT NO.: D00447427| JNJ DEPUY ORTHOPEDIC LOT NO.: 319786| IMPLANT NAME: SCREW CANC PINNACLE 6.5X20MM -| MANUFACTURER: JNJ DEPUY ORTHOPEDICS, |