EVIS EXERA III COLONOVIDEOSCOPE
Report
- Report Number
- 9610595-2025-05222
- Event Type
- Malfunction
- Date Received
- April 3, 2025
- Date of Event
- March 14, 2025
- Report Date
- April 3, 2025
- Manufacturer
- AIZU OLYMPUS CO., LTD.
- Product Code
- FDF
- UDI-DI
- 04953170305115
- PMA / PMN Number
- K131780
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- 003
Narratives
THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: 1- AW-CYLINDER (TUBE) HAS FOREIGN OBJECTS. (WHITE FOREIGN MATERIAL). 2-AW-TUBE HAS FOREIGN OBJECTS. (WHITE FOREIGN MATERIAL). THE WHITE FOREIGN MATERIAL DETECTED BY INCOME INSPECTION HAS BEEN HANDLED BY OMSC CAPA-201632. ACCORDING TO PREVIOUS INVESTIGATIONS, THE USE OF PRODUCTS THAT CONTAIN SILICONE CAN CAUSE DEPOSITS OF WHITE FOREIGN MATERIAL. SILICONE IS FOR EXAMPLE INCLUDED IN SIMETHICONE, A DEFOAMING AGENT, LUBRICANTS AND SO ON. IF THE CUSTOMER USED THEM, THERE WAS RISK THAT FOREIGN MATERIAL WAS REMAINED IN THE CHANNEL EVEN IF THE REPROCESSING WAS CONDUCTED IN ACCORDANCE WITH IFU. SIMETHICONE AND PETROLEUM/OIL/SILICONE-BASED LUBRICANTS ARE NON-WATER SOLUBLE AND THUS NOT RECOMMENDED FOR USE BY OLYMPUS. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FLEX VIDEO SCOPE EXHIBITED A FOREING MATERIAL ISSUE. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 653985 | EVIS EXERA III COLONOVIDEOSCOPE | COLONOVIDEOSCOPE | FDF | AIZU OLYMPUS CO., LTD. | CF-HQ190L | 04953170305115 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |