EVIS EXERA III COLONOVIDEOSCOPE

MDR report key: 21767728 · Received April 3, 2025

Report

Report Number
9610595-2025-05222
Event Type
Malfunction
Date Received
April 3, 2025
Date of Event
March 14, 2025
Report Date
April 3, 2025
Manufacturer
AIZU OLYMPUS CO., LTD.
Product Code
FDF
UDI-DI
04953170305115
PMA / PMN Number
K131780
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: 1- AW-CYLINDER (TUBE) HAS FOREIGN OBJECTS. (WHITE FOREIGN MATERIAL). 2-AW-TUBE HAS FOREIGN OBJECTS. (WHITE FOREIGN MATERIAL). THE WHITE FOREIGN MATERIAL DETECTED BY INCOME INSPECTION HAS BEEN HANDLED BY OMSC CAPA-201632. ACCORDING TO PREVIOUS INVESTIGATIONS, THE USE OF PRODUCTS THAT CONTAIN SILICONE CAN CAUSE DEPOSITS OF WHITE FOREIGN MATERIAL. SILICONE IS FOR EXAMPLE INCLUDED IN SIMETHICONE, A DEFOAMING AGENT, LUBRICANTS AND SO ON. IF THE CUSTOMER USED THEM, THERE WAS RISK THAT FOREIGN MATERIAL WAS REMAINED IN THE CHANNEL EVEN IF THE REPROCESSING WAS CONDUCTED IN ACCORDANCE WITH IFU. SIMETHICONE AND PETROLEUM/OIL/SILICONE-BASED LUBRICANTS ARE NON-WATER SOLUBLE AND THUS NOT RECOMMENDED FOR USE BY OLYMPUS. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FLEX VIDEO SCOPE EXHIBITED A FOREING MATERIAL ISSUE. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
653985 EVIS EXERA III COLONOVIDEOSCOPE COLONOVIDEOSCOPE FDF AIZU OLYMPUS CO., LTD. CF-HQ190L 04953170305115

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown