FDA Adverse Event Other Summary report: N

PASSPORT TROCAR

MDR report key: 2176771 · Received July 14, 2011

Report

Report Number
1651380-2011-00001
Event Type
Other
Date Received
July 14, 2011
Date of Event
June 9, 2011
Report Date
July 12, 2011
Manufacturer
PATTON SURGICAL CORP.
Product Code
GCJ
PMA / PMN Number
K080161
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
RISK MANAGER

Narratives

Additional Manufacturer Narrative · 1

ALL UNITS TESTED WERE IN ORIGINAL, SINGLE UNIT, UNOPENED PACKAGES. EACH CANNULA TUBE ASSEMBLY WAS SECURED INTO A TEXT FIXTURE, AT 30 DEGREES RELATIVE TO VERTICAL, AND FORCE WAS APPLIED AT A RATE OF 12 IN/MIN TO THREE LOCATIONS ALONG THE CANNULA TUBE; PROXIMAL, CENTER AND DISTAL USING A CHARTILLON PCM 2001 AND BG100RT LOAD CELL. NO UNITS SHATTERED UNDER THE DYNAMIC LOAD APPLIED, AT EITHER THE PROXIMAL, CENTER OR DISTAL LOCATIONS EVALUATED. DETAILED TEST DATA IS ON FILE AT PATTON SURGICAL CORPORATION.

Description of Event or Problem · 1

WHEN THE THORASCOPIC CAMERA PORT WAS REMOVED, STAFF OBSERVED THAT THE LAPAROSCOPIC TROCAR SHATTERED. SURGEON WAS ABLE TO RETRIEVE 2 BROKEN PIECES OF THE PORT FROM THE THORACIC CAVITY; UNABLE TO FIND 3RD PIECE (APPROX 5MM X 1CM). LEFT HEMITHORAX WAS THOROUGHLY INSPECTED AND PIECE NOT FOUND. CHEST X-RAY NEGATIVE. PIECE NOT RADIO-OPAQUE. PIECE NOT LOCATED AFTER SEARCH IN OPERATING ROOM. SURGEON EXPLAINED IF PIECE RETAINED, NO HARM TO PATIENT. ORIGINAL INTENDED PROCEDURE: FLEXIBLE BRONCHOSCOPY, ENDOBRONCHIAL ULTRASOUND, TRANSBRONCHIAL NEEDLE ASPIRATION, MEDIASTINOSCOPY, VIDEO ASSISTED THORASCOPY, LEFT LOWER LOBE LOBECTOMY, LYMPH NODE DISSECTION, INTERCOSTAL NERVE BLOCK FOR LUNG CA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PASSPORT TROCAR PLASTIC TROCAR GCJ PATTON SURGICAL CORP. 12100ES-11 8096-01

Patients

Seq Age Sex Outcome Treatment
1 73 YR