FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER

MDR report key: 2176393 · Received July 25, 2011

Report

Report Number
2050012-2011-03205
Event Type
Malfunction
Date Received
July 25, 2011
Date of Event
June 19, 2011
Report Date
June 20, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ON (B)(6) 2011 A BEC FIELD SERVICE ENGINEER (FSE) FOUND A POOR SIGNAL FROM THE WASH SOLUTION CANISTER FLOAT SWITCH CAUSING THE CANISTER TO OVERFILL AND LEAK OUT REGULATOR. FSE REPLACED THE FLOAT SWITCH FOR WASH SOLUTION AND VERIFIED OPERATION. REPAIR VERIFIED PER ESTABLISHED PROCEDURES. BEC INTERNAL NOTIFICATION NUMBER IS (B)(6).

Description of Event or Problem · 1

ON (B)(6) 2011, A CUSTOMER CONTACTED BECKMAN COULTER INC (BEC) CUSTOMER TECHNICAL SUPPORT (CTS) TO REPORT A LEAK FROM THE 10 PSI AND AIR PRESSURE WAS RUNNING LOW. ON (B)(6) 2011, THE CUSTOMER STATED THEY WERE RECEIVING AIR PRESSURE HIGH ERROR AND CUVETTES RAILING WATER BLANK. CUSTOMER FOUND WASH SOLUTION EMPTY AND TRIED FILLING CANISTERS IN PRIME AND FOAMY LIQUID SPEWED OUT OF THE 10 PSI REGULATOR. NO INJURY OR EXPOSURE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER FOR CLINICAL USE JJE BECKMAN COULTER INC. DXC 600 PRO N/A