FDA Adverse Event Malfunction Summary report: N

INFINITY ACUTE CARE SYSTEM (M540)

MDR report key: 21762175 · Received April 3, 2025

Report

Report Number
1220063-2025-00027
Event Type
Malfunction
Date Received
April 3, 2025
Date of Event
March 13, 2025
Report Date
September 2, 2025
Manufacturer
DRAEGER MEDICAL SYSTEMS, INC
Product Code
MHX
UDI-DI
04049098054454
PMA / PMN Number
K113798
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION HAS STARTED. A FOLLOW UP REPORT WILL BE SUBMITTED UPON COMPLETION.

Additional Manufacturer Narrative · 0

A DRAEGER TECHNICIAN WENT TO THE FACILITY TO RETRIEVE THE COCKPIT AND THE M540 LOGS FOR THE DATE AND TIME OF EVENT. THE LOG ANALYSIS CONFIRMED THE DISCONNECT, BUT THE EXACT ROOT CAUSE COULD NOT BE IDENTIFIED. THE ENTIRE IACS SYSTEM WAS REQUESTED WITH BOTH THE C500 AND THE M540 FOR ANALYSIS. THE IACS SYSTEM WAS CONNECTED TO THE DRAEGER INTERNAL NETWORK FOR TESTING AND NO OBSERVABLE DISCONNECT OCCURRED IN THE TWO MONTHS RUNNING AN ECG AND WAVEFORM SIMULATION. THE CUSTOMER WAS SENT A REPLACEMENT IACS SYSTEM WHICH WAS CONFIRMED TO BE WORKING ACCORDING TO MANUFACTURER'S SPECIFICATIONS.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE INFINITY M540 MONITOR DISCONNECTED FROM THE INFINITY MEDICAL COCKPIT C500 DURING A PATIENT CARDIAC PROCEDURE AND THE M540 MONITOR WAS SWAPPED OUT AS A RESULT. ALTHOUGH AN INTERRUPTION IN PATIENT MONITORING OCCURRED, NO ADVERSE PATIENT IMPACT WAS REPORTED.

Description of Event or Problem · 0

IT WAS REPORTED THAT THE INFINITY M540 MONITOR DISCONNECTED FROM THE INFINITY MEDICAL COCKPIT C500 DURING A PATIENT CARDIAC PROCEDURE AND THE M540 MONITOR WAS SWAPPED OUT AS A RESULT. ALTHOUGH AN INTERRUPTION IN PATIENT MONITORING OCCURRED, NO ADVERSE PATIENT IMPACT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
731229 INFINITY ACUTE CARE SYSTEM (M540) PHYSIOLOGICAL MONITORING SYSTEMS MHX DRAEGER MEDICAL SYSTEMS, INC N/A N/A 04049098054454

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown