PENTAX
Report
- Report Number
- 9610877-2025-51518
- Event Type
- Malfunction
- Date Received
- April 3, 2025
- Date of Event
- March 18, 2025
- Report Date
- April 3, 2025
- Manufacturer
- HOYA CORPORATION PENTAX TOKYO OFFICE
- Product Code
- EQL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
G4: THIS DEVICE IS CLASS1 PRODUCT SO THAT 510K# IS BLANK. THE PRODUCT WAS RETURNED TO PENTAX MEDICAL FOR REPAIR. OUR TECHNICIAN CHECKED THE RETURNED UNIT AND CONFIRMED THAT THE CCD MODULE WITH DRIVE PCB CLOUDY (NOT CLEAR VIEW). BASED ON THE RESULT, WE CONCLUDED THAT IT WAS CAUSED DUE TO THE MOISTURE CONDENSATION IN THE CCD MODULE WITH DRIVE PCB. IN ADDITION, OUR TECHNICIAN CONFIRMED THAT THE U/D KNOB BROKEN, THE INSERTION FLEXIBLE TUBE BUCKLED, THE LIGHT GUIDE CABLE FOR CONTROL BODY BUCKLED, THE REMOTE CONTROL BUTTONS CRACKED, THE REMOTE CONTROL BUTTONS LEAK, AND THE LCB (LIGHT CARRYING BUNDLE) DEFECTIVE; HOWEVER, THESE DEFECTS ARE NOT THE MAIN CAUSE, AND/OR IRRELEVANT TO THE ALLEGED COMPLAINT. BASED ON THE TECHNICAL REPORT ""HR-RPT-0586(IMAGE FAILURE)"" AND/OR THE RISK ANALYSIS RESULTS, IT WAS EVALUATED TO SUBMIT MDR.
THE TIME OF EVENT IS UNKNOWN. THERE WAS NO REPORT OF PATIENT HARM. VIDEO IMAGE FAILURE (CLOUDY).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 732060 | PENTAX | VIDEO LARYNGOSTROBOSCOPE (90 SERIES) | EQL | HOYA CORPORATION PENTAX TOKYO OFFICE | VLS-1190STK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |