CPS LOCATOR 3D DELIVERY CATHETER
Report
- Report Number
- 3015970743-2025-00004
- Event Type
- Injury
- Date Received
- April 2, 2025
- Date of Event
- March 3, 2025
- Report Date
- April 2, 2025
- Manufacturer
- CENTERPOINT SYSTEMS LLC
- Product Code
- DQY
- PMA / PMN Number
- K230363
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
UPON RECEIPT OF THE COMPLAINT INFORMATION, CENTERPOINT INQUIRED FOR ADDITIONAL INFORMATION AS THERE DID NOT APPEAR TO BE A DEVICE PROBLEM WHICH CONTRIBUTED TO THE SEPTUM PERFORATION DURING THIS EVENT. CENTERPOINT CONFIRMED FROM THE DISTRIBUTOR THAT THE PHYSICIAN INDICATED THAT THE SEPTUM OF THE PATIENT WAS VERY THIN, AND THE PERFORATION WAS CAUSED BY THE LEAD ONLY. DIAGNOSTIC IMAGING WAS PERFORMED TO CONFIRM THE PERFORATION, BUT IT WAS INCONCLUSIVE, THEREFORE THE SEPTUM PERFORATION WAS ASSUMED BY THE LEAD IMPEDENCE VALUES. THE PATIENT REMAINED STABLE THROUGHOUT THE ENTIRE PROCEDURE, AND NO SERIOUS HEALTH CONSEQUENCE RESULTED FROM THIS EVENT. ALTHOUGH IT WAS CLEAR FROM THE PHYSICIAN AND THE DISTRIBUTOR THAT THERE WAS NO FAULT IN THE CENTERPOINT PRODUCT (CPS LOCATOR 3D DELIVERY CATHETER), AND THE PERFORATION WAS RELATED TO THE LEAD ONLY AND DID NOT RESULT IN AN ADVERSE EVENT, CENTERPOINT IS REPORTING THIS EVENT OUT OF CAUTION TO COMPLY WITH APPLICABLE REGULATORY REQUIREMENTS.
DURING CONDUCTION SYSTEM PACING PHYSICIAN WAS PLACING THE LEAD INTO THE LEFT BUNDLE BRANCH THROUGH CPS LOCATOR 3D CATHETER. DURING THE POSITIONING OF THE LEAD, IT WAS OBSERVED THAT THE LEAD IMPEDENCE WAS VERY HIGH AT POINT WHICH IS INDICATIVE OF PERFORATION. THEN THE LEAD WAS PULLED OUT WASHED THOROUGHLY AND AGAIN POSITITIONED. LIKE FIRST INSTANCE, AGAIN IT PERFORATED THE SEPTUM. PHYSICIAN THEN DECIDED TO PERFORM CONVENTIONAL DUAL CHAMBER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2371767 | CPS LOCATOR 3D DELIVERY CATHETER | PERCUATNEOUS CATHETER | DQY | CENTERPOINT SYSTEMS LLC | DS2C200-42 | CL12490 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |