FDA Adverse Event Malfunction Summary report: N

BD SYRINGE 10ML LL TIP BULK CONVENIENCE PAK

MDR report key: 21757776 · Received April 2, 2025

Report

Report Number
1213809-2025-00228
Event Type
Malfunction
Date Received
April 2, 2025
Date of Event
February 17, 2025
Report Date
April 21, 2025
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903096054
PMA / PMN Number
K110771
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PR (B)(4) - SUPPLEMENTAL MDR - FOREIGN MATTER. ADDITIONAL INFORMATION: TWO POSSIBLE LOT PROVIDED LOT 4073231 AND LOT 4192355 EVALUATION: TWO PHOTO AND ONE SAMPLE WERE PROVIDED TO OUR QUALITY TEAM FOR INVESTIGATION. UPON EVALUATION, IT WAS OBSERVED THAT A FULLY ASSEMBLED SYRINGE FILLED WITH TRANSPARENT FLUID, CONTAINING A LARGE INSECT FLOATING IN THE FLUID PATH INSIDE THE BARREL. THE OBSERVED CONDITION IS NON-CONFORMING TO PRODUCT SPECIFICATIONS. THE POTENTIAL ROOT CAUSE FOR THE FOREIGN MATTER IN FLUID PATH DEFECT IS ASSOCIATED WITH THE ASSEMBLY PROCESS. AS CORRECTIVE ACTION, QUALITY ALERT WILL BE SENT TO THE PLANT. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBERS 4073231 AND 4192355. A REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION WAS PROVIDED

Description of Event or Problem · 0

MATERIAL#: 309605, BATCH#: UNKNOWN. IT WAS REPORTED THAT THE BD SYRINGE 10ML LL TIP BULK CONVENIENCE PAK HAD FOREIGN MATTER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: RCC RECEIVED A COMPLAINT VIA EMAIL. DURING VISUAL INSPECTION OF FINISHED PRODUCT, AN INSECT WAS VISUALIZED INSIDE THE SYRINGE FLOATING IN THE SOLUTION. ITEM NUMBER: 309605, LOT NUMBER: 2000205.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
634206 BD SYRINGE 10ML LL TIP BULK CONVENIENCE PAK PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS UNKNOWN 30382903096054

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown