FDA Adverse Event Malfunction Summary report: N

BD FACSLYRIC¿

MDR report key: 21756073 · Received April 2, 2025

Report

Report Number
3027394506-2025-00012
Event Type
Malfunction
Date Received
April 2, 2025
Date of Event
March 5, 2025
Report Date
June 26, 2025
Manufacturer
BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES
Product Code
OYE
UDI-DI
00382906635181
PMA / PMN Number
BK230835
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.6 INVESTIGATION SUMMARY: BASED ON THE INVESTIGATION RESULTS, THE REPORTED ISSUE OF THE CUSTOMER EXPERIENCING CONTAMINATION AND CARRYOVER WAS CONFIRMED. INVESTIGATION RESULTS THAT WERE PERFORMED ON THE INDICATED FAILURE MODE WERE THE FOLLOWING: THE COMPLAINT TREND AND SERVICE NOTES WERE REVIEWED. THE POTENTIAL CAUSE FOR THE CUSTOMER EXPERIENCING CONTAMINATION AND CARRYOVER WAS DETERMINED TO BE LINKED TO A DEFECTIVE PUMP IN THE FLUIDIC SYSTEM. THE ISSUE WAS RESOLVED AFTER THE PART WAS REPLACED. NO FURTHER PROBLEMS WERE NOTED, AND THE INSTRUMENT WAS CONFIRMED TO BE FUNCTIONING AS EXPECTED.

Additional Manufacturer Narrative · 0

ADDITIONAL PMA/510K CODES ARE: K170974, K201814. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 0

IT WAS REPORTED THAT CARRYOVER BETWEEN PATIENT SAMPLES WAS OBSERVED AFTER USAGE OF BD FACSLYRIC¿ . THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: SPOKE WITH CUSTOMER AND THEY ARE GETTING LASER DELAY SETUP ERRORS DURING PQC. THEY HAVE COMPLETED A FEW LONG CLEANS IN THE LAST FEW WEEKS BECAUSE THEY ARE STILL HAVING SOME CARRYOVER ISSUES BETWEEN SAMPLES. SYSTEM HAS BEEN HAVING ISSUES WITH CARRYOVER, CLEANING THE SYSTEM CLEARS UP THE ISSUE BUT IT COMES BACK AFTER A FEW DAYS. NO RESULTS WERE USED IF CARRYOVER WAS PRESENT.

Description of Event or Problem · 0

IT WAS REPORTED THAT CARRYOVER BETWEEN PATIENT SAMPLES WAS OBSERVED AFTER USAGE OF BD FACSLYRIC¿ . THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: SPOKE WITH CUSTOMER AND THEY ARE GETTING LASER DELAY SETUP ERRORS DURING PQC. THEY HAVE COMPLETED A FEW LONG CLEANS IN THE LAST FEW WEEKS BECAUSE THEY ARE STILL HAVING SOME CARRYOVER ISSUES BETWEEN SAMPLES. SYSTEM HAS BEEN HAVING ISSUES WITH CARRYOVER, CLEANING THE SYSTEM CLEARS UP THE ISSUE BUT IT COMES BACK AFTER A FEW DAYS. NO RESULTS WERE USED IF CARRYOVER WAS PRESENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2372642 BD FACSLYRIC¿ FLOW CYTOMETRIC REAGENTS AND ACCESS OYE BECTON, DICKINSON AND COMPANY, BD BIOSCIENCES 00382906635181

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown