4C -ES CELL CONTROL

MDR report key: 2175321 · Received June 15, 2011

Report

Report Number
1061932-2011-00571
Event Type
Malfunction
Date Received
June 15, 2011
Date of Event
October 13, 2008
Report Date
October 13, 2008
Manufacturer
BECKMAN COULTER, INC.
Product Code
JPK
PMA / PMN Number
K010064
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

BECKMAN COULTER INC. PRODUCT LABELING STATES THAT PPE MUST BE WORN WHEN HANDLING THE PRODUCT. SPECIFICALLY, LABELING STATES: THIS SPECIMEN/REAGENT SHOULD BE HANDLED AT BIOSAFETY LEVEL 2, AS RECOMMENDED FOR ANY POTENTIALLY INFECTIOUS HUMAN SERUM OR BLOOD SPECIMEN IN THE (B)(6) MANUAL "BIOSAFETY IN MICROBIOLOGICAL AND BIOMEDICAL LABORATORIES," 1988. CUSTOMER CANCELED A PLANNED FIELD SERVICE VISIT TO EVALUATE THE ANALYZER BECAUSE THE OTHER CONTROL VIALS DID NOT LEAK. THE PRODUCT WAS USED FOR APPROX 15 PIERCES BEFORE IT LEAKED. THE OPEN VIAL CLAIM IS THAT THE PROCEDURE LISTED IN THE INSTRUCTIONS FOR USE CAN BE PERFORMED A MAXIMUM OF 20 TIMES WITHIN 35 DAYS. ROOT CAUSE FOR THE FAILURE OF THE VIAL STOPPER WAS NOT DETERMINED. THIS REPORTABLE EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW CONDUCTED BETWEEN (B)(4) 2008 AND (B)(4) 2010 OF COMPLAINTS FOR ADD'L REPORTABLE EVENTS.

Description of Event or Problem · 1

CUSTOMER REPORTED A POTENTIAL BIOHAZARD EVENT WHEN A NORMAL CONTROL VIAL OF 4C -ES CELL CONTROL LEAKED FROM THE CAP WHEN USING A COULTER AC-T DIFF 2 ANALYZER ON (B)(6) 2008. THE OPERATOR WAS NOT WEARING PERSONAL PROTECTIVE EQUIPMENT. THE OPERATOR WAS EXPOSED TO THE CONTROL MATERIAL, HOWEVER, THERE WAS NO EXPOSURE TO MUCOUS MEMBRANES OR OPEN LESIONS. THE OPERATOR DID NOT SEEK MEDICAL ATTENTION. NO REPORTS OF DEATH OR SERIOUS INJURY, AND NO AFFECT TO OPERATOR SAFETY AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 4C -ES CELL CONTROL JPK BECKMAN COULTER, INC. NA 079200

Patients

Seq Age Sex Outcome Treatment
1 UNK COULTER AC-T DIFF 2 ANALYZER