FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER

MDR report key: 2174110 · Received July 24, 2011

Report

Report Number
2050012-2011-03204
Event Type
Malfunction
Date Received
July 24, 2011
Date of Event
June 19, 2011
Report Date
June 19, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER PRIMED AND CALIBRATED THE INSTRUMENT AND QC RESULTS WERE WITHIN RANGE. THE CUSTOMER REVIEWED STATUS SCREEN, NO ISSUES WERE NOTED AND NO LEAKS. THE CUSTOMER REQUESTED A SERVICE VISIT TO CHECK OPERATION OF THE INSTRUMENT. ON (B)(4) 2011, A BEC FIELD SERVICE ENGINEER (FSE) PERFORMED SERVICE ON THE INSTRUMENT. FSE VERIFIED THE LEAK ISSUE AND NOTED THE WASH SOLUTION CANISTER WAS OVERFILLED. FSE REPLACED THE WASH SOLUTION CANISTER FLOAT SWITCH FSE REPLACED VALVES V12 AND V14 (BOTH ASSOCIATED WITH THE WASH SOLUTION CANISTER). FSE PRIMED THE HYDRO AND FILLED ALL CANISTERS AND NO FURTHER ISSUES OBSERVED. REPAIRS VERIFIED PER ESTABLISHED PROCEDURES. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC (BEC) CUSTOMER TECHNICAL SUPPORT (CTS) TO REPORT A LIQUID LEAK OUT OF THE MIDDLE OF THE INSTRUMENT AND ONTO THE FLOOR. THE CUSTOMER STATED THE FLUID WAS A LIGHT YELLOW. CREATININE REAGENT CONTAINER WAS CHECKED FOR DAMAGES AND NONE WERE NOTED. NO INJURY OR EXPOSURE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC SYNCHRON ® CHEMISTRY ANALYZER DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER FOR CLINICAL USE JJE BECKMAN COULTER INC. DXC 800 PRO N/A

Patients

Seq Age Sex Outcome Treatment
1