DA VINCI SI SURGICAL SYSTEM
Report
- Report Number
- 2955842-2011-00231
- Event Type
- Other
- Date Received
- July 24, 2011
- Date of Event
- June 20, 2011
- Report Date
- June 20, 2011
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K081137
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MT
- Reporter Occupation
- NURSE
Narratives
THE INVESTIGATION CONDUCTED BY FIELD SERVICE ENGINEERING FOUND SYSTEM ERROR CODE 23017 AND 25588 IN THE SYSTEM ERROR LOGS. IT WAS DETERMINED THAT THESE ERRORS WERE ASSOCIATED WITH THE ENDOSCOPIC CAMERA MANIPULATOR (ECM) ARM. THE ECM IS THE CAMERA ARM LOCATED ON THE PATIENT SIDE CART WHICH PROVIDES THE STERILE INTERFACE FOR THE 3D ENDOSCOPE. THE SYSTEM WAS REPAIRED BY REPLACING THE AFFECTED ECM. THE SYSTEM ALARM (SYSTEM GENERATED FAULT CODE) FUNCTIONED AS DESIGNED AND THERE WAS NO INJURY TO THE PATIENT. SYSTEM ERROR CODE 23017 APPEARS WHEN THE DA VINCI S SAFETY SYSTEM DETERMINES THAT A MOTOR DID NOT RESPOND AS EXPECTED AND THE MEASURED MOTION DID NOT MATCH THE INTERNAL SIMULATION OF THE MOTOR. UPON DETERMINING THIS CONDITION, THE SAFETY SYSTEMS PUT DA VINCI IN A RECOVERABLE SAFE STATE. SYSTEM ERROR CODE 25588 APPEARS WHEN THE DA VINCI S SAFETY SYSTEM DETERMINES THAT A FLAT FLEX CABLE CONNECTING TO THE EMBEDDED SERIALIZER IN MASTER PLATFORM (ESPM) BOARD WAS NOT FULLY INSERTED. UPON DETERMINING THIS CONDITION, THE SAFETY SYSTEMS PUT DA VINCI IN A NON-RECOVERABLE SAFE STATE. THE ECM WAS RETURNED TO ISI FOR FAILURE ANALYSIS INVESTIGATION. ENGINEERING CONFIRMED THE CUSTOMER REPORTED PROBLEM AND REPLACED THE MOTOR ENCODER AND FULLY INSERTED THE FLAT FLEX CABLE INTO THE ESPM TO REPAIR THE ECM. AS OF JULY 18, 2011, THERE HAVE BEEN NO REPORTED RECURRENCES OF THE ISSUE AT THIS HOSPITAL.
IT WAS REPORTED THAT DURING THE BEGINNING OF A DA VINCI SI PROSTATECTOMY PROCEDURE, THE SITE EXPERIENCED ERROR CODE 23017 AND THE ENDOSCOPIC CAMERA MANIPULATOR'S(ECM) LED WAS ILLUMINATED YELLOW. WITH THE ASSISTANCE OF A ISI TECHNICAL SUPPORT ENGINEER (TSE), THE SITE ATTEMPTED TO TROUBLESHOOT THE ISSUE BY APPLYING EMERGENCY POWER OFF (EPO); HOWEVER, NON RECOVERABLE SYSTEM ERROR CODE 25588 OCCURRED. AT THIS TIME, THE DECISION WAS MADE TO PERFORM THE SURGERY USING TRADITIONAL OPEN TECHNIQUES. THERE WAS NO PATIENT INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | DA VINCI SI SURGICAL SYSTEM | ENDOSCOPIC INSTRUMENT CONTROL SYSTEM | NAY | INTUITIVE SURGICAL,INC. | IS3000 A6.0P6 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other | DA VINCI SI SYSTEM INSTRUMENT, ACCESSORIES & ESU |