FDA Adverse Event Malfunction Summary report: N

PROMUS ELEMENT EVEROLIMUS-ELUTING CORONARY STENT SYSTEM

MDR report key: 2173919 · Received July 24, 2011

Report

Report Number
2134265-2011-02921
Event Type
Malfunction
Date Received
July 24, 2011
Date of Event
May 16, 2011
Report Date
June 30, 2011
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
SIMILAR
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE IS COMBINATION PRODUCT. DEVICE EVALUATED BY MFR: A VISUAL AND MICROSCOPIC EXAMINATION OF THE DEVICE IDENTIFIED DISTAL STENT DAMAGE. STRUTS ON ROWS THE 1ST AND 2ND ROWS FROM THE DISTAL END WERE SLIGHTLY RAISED AND MISALIGNED. THE DISTAL END OF THE STENT WAS RESTING ON THE PROXIMAL END OF THE DISTAL MARKERBAND. THE OUTER DIAMETER (OD) OF THE DAMAGED SECTION WAS MEASURED AT 1.32MM. THE OD OF THE STENT AREA WHERE NO DAMAGE WAS PRESENT WAS MEASURED AT 1.18 MM. NO ISSUES WERE NOTED WITH THE BALLOON AND TIP SECTIONS OF THE DEVICE. THE BALLOON WAS TIGHTLY WRAPPED AND WAS NOT SUBJECTED TO POSITIVE PRESSURE. NO KINKS OR DAMAGE WERE NOTICED ALONG THE SHAFT OF THE DEVICE. A 0.015 INCH PRODUCT MANDREL WAS INSERTED THROUGH THE LUMEN WITH NO RESTRICTIONS NOTED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

REPORTABLE BASED ON ANALYSIS COMPLETED ON (B)(6) 2011. IT WAS REPORTED THAT DURING A STENTING TREATMENT PROCEDURE, DIFFICULTY CROSSING OCCURRED. THE 84% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING CORONARY ARTERY (LAD). THE LESION WAS PREDILATED AND A 3.0X16MM PROMUS ELEMENT STENT DELIVERY SYSTEM WAS ADVANCED BUT WAS UNABLE TO CROSS THE LESION. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS STABLE. HOWEVER, PRODUCT ANALYSIS REVEALED STENT DAMAGE. THIS PRODUCT IS ONLY OUS APPROVED BUT IT IS SIMILAR TO AN APPROVED US DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROMUS ELEMENT EVEROLIMUS-ELUTING CORONARY STENT SYSTEM STENT, CORONARY, DRUG-ELUTING NIQ BOSTON SCIENTIFIC - GALWAY H7493911316300 13190509

Patients

Seq Age Sex Outcome Treatment
1 58 YR