PROMUS ELEMENT EVEROLIMUS-ELUTING CORONARY STENT SYSTEM
Report
- Report Number
- 2134265-2011-02921
- Event Type
- Malfunction
- Date Received
- July 24, 2011
- Date of Event
- May 16, 2011
- Report Date
- June 30, 2011
- Manufacturer
- BOSTON SCIENTIFIC - GALWAY
- Product Code
- NIQ
- PMA / PMN Number
- SIMILAR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KR
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE IS COMBINATION PRODUCT. DEVICE EVALUATED BY MFR: A VISUAL AND MICROSCOPIC EXAMINATION OF THE DEVICE IDENTIFIED DISTAL STENT DAMAGE. STRUTS ON ROWS THE 1ST AND 2ND ROWS FROM THE DISTAL END WERE SLIGHTLY RAISED AND MISALIGNED. THE DISTAL END OF THE STENT WAS RESTING ON THE PROXIMAL END OF THE DISTAL MARKERBAND. THE OUTER DIAMETER (OD) OF THE DAMAGED SECTION WAS MEASURED AT 1.32MM. THE OD OF THE STENT AREA WHERE NO DAMAGE WAS PRESENT WAS MEASURED AT 1.18 MM. NO ISSUES WERE NOTED WITH THE BALLOON AND TIP SECTIONS OF THE DEVICE. THE BALLOON WAS TIGHTLY WRAPPED AND WAS NOT SUBJECTED TO POSITIVE PRESSURE. NO KINKS OR DAMAGE WERE NOTICED ALONG THE SHAFT OF THE DEVICE. A 0.015 INCH PRODUCT MANDREL WAS INSERTED THROUGH THE LUMEN WITH NO RESTRICTIONS NOTED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).
REPORTABLE BASED ON ANALYSIS COMPLETED ON (B)(6) 2011. IT WAS REPORTED THAT DURING A STENTING TREATMENT PROCEDURE, DIFFICULTY CROSSING OCCURRED. THE 84% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING CORONARY ARTERY (LAD). THE LESION WAS PREDILATED AND A 3.0X16MM PROMUS ELEMENT STENT DELIVERY SYSTEM WAS ADVANCED BUT WAS UNABLE TO CROSS THE LESION. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS STABLE. HOWEVER, PRODUCT ANALYSIS REVEALED STENT DAMAGE. THIS PRODUCT IS ONLY OUS APPROVED BUT IT IS SIMILAR TO AN APPROVED US DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PROMUS ELEMENT EVEROLIMUS-ELUTING CORONARY STENT SYSTEM | STENT, CORONARY, DRUG-ELUTING | NIQ | BOSTON SCIENTIFIC - GALWAY | H7493911316300 | 13190509 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR |