FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2173894 · Received July 24, 2011

Report

Report Number
6000001-2011-13655
Event Type
Malfunction
Date Received
July 24, 2011
Date of Event
June 28, 2011
Report Date
June 28, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K041191
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH AN INOPERATIVE "SELECT" BUTTON ON THE CHANNEL A KEYPAD WAS CONFIRMED DURING PRODUCT EVALUATION. THIS CONDITION WAS CAUSED BY FLUID INTRUSION. THE PUMP HEAD MODULE (PHM) KEYPAD WAS REPLACED TO CORRECT THE REPORTED CONDITION. ADDITIONAL INFORMATION: A SERVICE HISTORY REVIEW REVEALED THAT THIS DEVICE WAS PREVIOUSLY SERVICED FOR THE REPORTED CONDITION. DURING PREVIOUS SERVICE EVENTS, THE PHM KEYPAD WAS REPLACED. THIS INVOLVED A COLLEAGUE P1.5 INFUSION PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.48.92. THIS ISSUE HAS BEEN ESCALATED TO CAPA.

Description of Event or Problem · 1

BAXTER (B)(4) RECEIVED A REPORT OF A COLLEAGUE INFUSION PUMP WITH INOPERATIVE "SELECT" BUTTON ON THE CHANNEL A KEYPAD. IT IS UNKNOWN WHEN OR IN WHICH CARE AREA THIS EVENT OCCURRED. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THERE WAS NO REPORT OF PATIENT INVOLVEMENT, PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS PUMP IS CURRENTLY UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PRE P1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1