TRANSSEPTAL NEEDLE, BRK SERIES
Report
- Report Number
- 3008452825-2025-00145
- Event Type
- Injury
- Date Received
- April 1, 2025
- Report Date
- April 11, 2025
- Manufacturer
- ST. JUDE MEDICAL
- Product Code
- DRC
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TU
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
ADDITIONAL INFORMATION: B5, G3, H2, H6, H11. THE BROCKENBROUGH NEEDLE / SHEATH DID NOT CONTRIBUTE DIRECTLY TO THESE COMPLICATIONS. THESE EVENTS WERE DUE TO TECHNICAL CHALLENGES AND COMPLEX CARDIAC ANATOMY.
THE RESULTS OF THE INVESTIGATION ARE INCONCLUSIVE SINCE THE DEVICE WAS NOT RETURNED FOR ANALYSIS. REVIEW OF THE DEVICE HISTORY RECORD WAS NOT POSSIBLE AS THE LOT NUMBER IS UNKNOWN. BASED ON THE INFORMATION RECEIVED, THE CAUSE OF THE REPORTED TAMPONADE / PERFORATION REMAINS UNKNOWN. PER THE IFU, CARDIAC PERFORATION IS A KNOWN RISK DURING THE USE OF THIS DEVICE.
THE BROCKENBROUGH NEEDLE / SHEATH DID NOT CONTRIBUTE DIRECTLY TO THESE COMPLICATIONS. THESE EVENTS WERE DUE TO TECHNICAL CHALLENGES AND COMPLEX CARDIAC ANATOMY.
THE FOLLOWING WAS PUBLISHED IN: PACING AND CLINICAL ELECTROPHYSIOLOGY "TRANSSEPTAL PUNCTURE IN CHILDREN WEIGHING LESS THAN 20 KG IN INVASIVE CARDIAC CATHETERIZATION AND ELECTROPHYSIOLOGY "; MERVE MAZE AYDEMI. THIS RETROSPECTIVE STUDY ANALYZED TSP PROCEDURES IN INFANTS AND CHILDREN <20 KG BETWEEN DECEMBER 2015 AND JANUARY 2023. TSPS WERE PERFORMED WITH A BIPLANE ANGIOGRAPHY SYSTEM IN THE CATHETER ANGIOGRAPHY LABORATORY. A BROCKENBROUGH NEEDLE (BRK TRANSSEPTAL NEEDLE; ABBOTT/ST. JUDE MEDICAL, INC.) WAS USED FOR TSP. IN PATIENTS WHOSE BROCKENBROUGH NEEDLES COULD NOT CROSS THE INTERATRIAL SEPTUM, TSP WAS PERFORMED BY APPLYING CAUTERY ENERGY OVER THE BROCKENBROUGH NEEDLE. NINETEEN PATIENTS WERE STUDIED (SEVEN FEMALES, MEDIAN AGE 2.3 YEARS, MEDIAN WEIGHT 10.3 KG), AND 63% REQUIRED TSP FOR CHD. THE CHD GROUP HAD YOUNGER PATIENTS (MEDIAN AGE: 6.25 MONTHS) WITH SMALLER WEIGHTS (MEDIAN: 5.3 KG) THAN THE CATHETER ABLATION GROUP. THERE WAS NO DIFFERENCE BETWEEN GROUPS IN TERMS OF GENDER AND SUCCESS RATES. PROCEDURE AND FLUOROSCOPY TIMES WERE SHORTER IN THE ABLATION GROUP (P < 0.05). AS PERICARDIAL TAMPONADE OCCURRED IN TWO PATIENTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1172086 | TRANSSEPTAL NEEDLE, BRK SERIES | TROCAR | DRC | ST. JUDE MEDICAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |