FDA Adverse Event Injury Summary report: N

INSIGNIA

MDR report key: 2173718 · Received July 23, 2011

Report

Report Number
2124215-2011-10911
Event Type
Injury
Date Received
July 23, 2011
Date of Event
June 21, 2011
Report Date
August 18, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NVZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. PIN GAUGE TESTING, DESIGNED TO VERIFY PROPER PORT DIMENSIONS, WAS COMPLETED. EACH PORT MEASURED AS EXPECTED. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Additional Manufacturer Narrative · 1

THE PRODUCT HAS NOT BEEN RETURNED. IF RETURNED, OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS EVENT WILL UPDATED.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR LEAD EXHIBITED IMPEDANCE MEASUREMENTS GREATER THAN 2000 OHMS. ADDITIONALLY, THERE WAS NOISE THAT WAS OVERSENSED RESULTING IN PACING INHIBITION. THE PHYSICIAN DISCONECTED THE DEVICE AND TESTED THE LEAD WHICH RESULTED IN NORMAL MEASUREMENTS. AS A RESULT, THE PHYSICIAN ELECTED TO REPLACE THE DEVICE ONLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INSIGNIA IMPLANTABLE PULSE GENERATOR NVZ GUIDANT CRM CLONMEL IRELAND 1198

Patients

Seq Age Sex Outcome Treatment
1 75 YR Life Threatening 1198| 4473