FDA Adverse Event Injury Summary report: N

VITALITY

MDR report key: 2173700 · Received July 23, 2011

Report

Report Number
2124215-2011-10943
Event Type
Injury
Date Received
July 23, 2011
Date of Event
June 18, 2011
Report Date
October 24, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL AVAILABLE INFORMATION INDICATES THAT THE DEVICE REMAINS IN SERVICE. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

ANALYSIS OF THE RETURNED PRODUCT IS CURRENTLY ONGOING. THIS EVENT WILL BE UPDATED UPON COMPLETION OF THE ANALYSIS.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, THE DEVICE WAS THOROUGHLY INSPECTED AND ANALYZED. THE DEVICE WAS THEN SUBJECTED TO A SERIES OF AUTOMATED DIAGNOSTIC TESTS THAT VERIFY THE PERFORMANCE OF THE DEFIBRILLATION, PACING SENSING, HIGH VOLTAGE SHOCKING AND RECORDING FUNCTIONS OF THE DEVICE. THE DEVICE PERFORMED NORMALLY THROUGHOUT ALL TESTS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE PATIENT IMPLANTED WITH THIS DEVICE WAS PRESENTED IN THE HOSPITAL WITH A VENTRICULAR TACHYCARDIA (VT) RATE OF 140. THE PATIENT WAS EXTERNALLY DEFIBRILLATED AS A RESULT OF THE DEVICE NOT BEING PROGRAMMED TO RECOGNIZE THAT RATE. THE DEVICE WAS REPROGRAMMED AND NO FURTHER OBSERVATIONS WERE REPORTED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED THAT THIS DEVICE WAS EXPLANTED FOR AN UNSPECIFIED REASON. THE DEVICE WAS RETURNED TO ALLOW FOR RELIABILITY ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VITALITY IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND T135

Patients

Seq Age Sex Outcome Treatment
1 68 YR Required Intervention 0145| T135| 1790