FDA Adverse Event Malfunction Summary report: N

FINELINE II

MDR report key: 2173687 · Received July 23, 2011

Report

Report Number
2124215-2011-10524
Event Type
Malfunction
Date Received
July 23, 2011
Date of Event
June 16, 2011
Report Date
August 5, 2011
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.

Additional Manufacturer Narrative · 1

SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS INVESTIGATION WILL BE UPDATED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING THE IMPLANT PROCEDURE HIGH OUT OF RANGE PACING IMPEDANCES WERE NOTED ON THIS RIGHT VENTRICULAR LEAD. THIS LEAD WAS NOT IMPLANTED AND WILL BE RETURNED FOR ANALYSIS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

--

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FINELINE II IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4480

Patients

Seq Age Sex Outcome Treatment
1