INSIGNIA
Report
- Report Number
- 2124215-2011-11938
- Event Type
- Injury
- Date Received
- July 23, 2011
- Date of Event
- June 21, 2011
- Report Date
- September 1, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NVZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE PRODUCT IS EXPECTED TO BE RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ANALYSIS.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, AN ENGINEERING LEVEL LONGEVITY PREDICTION CALCULATION WAS COMPLETED TO ASSESS THE RATE OF BATTERY DEPLETION. GIVEN THE PROGRAMMED PARAMETERS AND OTHER DATA STORED WITHIN THE MEMORY OF THE DEVICE, THE RESULTS OF THIS CALCULATION INDICATED THAT THE ACTUAL RATE OF BATTERY DEPLETION FELL WITHIN AN ACCEPTABLE RANGE. NEXT, A SERIES OF DIAGNOSTIC TESTS WERE CONDUCTED THAT VERIFIED THE PERFORMANCE OF PACING, SENSING, AND RECORDING FUNCTIONS. HAVING MET THE ENGINEERING LONGEVITY PREDICTION, FUNCTIONALLY PASSING ALL RETURNED PRODUCT TESTING, AND WITH NO FURTHER INFORMATION TO INDICATE A PRODUCT PERFORMANCE ISSUE, WE HAVE CONCLUDED THAT THIS DEVICE EXPERIENCED NORMAL BATTERY DEPLETION.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PATIENT WAS EXPERIENCING LIGHTHEADEDNESS. ELECTIVE REPLACEMENT TIME (ERT) WAS DECLARED AND LESS THAT A MONTH LATER END OF LIFE (EOL) WAS DECLARED. THIS DEVICE WAS EXPLANTED FOR NORMAL BATTERY DEPLETION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INSIGNIA | IMPLANTABLE PULSE GENERATOR | NVZ | GUIDANT CRM CLONMEL IRELAND | 1190 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR | Required Intervention | 1190| 4087 |