FDA Adverse Event Malfunction Summary report: N

COMMUNICATOR

MDR report key: 2173660 · Received July 23, 2011

Report

Report Number
2124215-2011-10667
Event Type
Malfunction
Date Received
July 23, 2011
Date of Event
June 20, 2011
Report Date
July 26, 2011
Manufacturer
EXTERNAL MANUFACTURER
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, THE COMMUNICATOR WAS RETURNED WITH BOTH THE POWER BRICK AND PHONE CORD. A VISUAL INSPECTION REVEALED NO PHYSICIAL DAMAGE. THE POWER BRICK DID NOT SUCCESSFULLY PASS THE VOLTAGE CHECK. FURTHER ANALYSIS REVEALED THE OUTSIDE CASE OF THE POWER BRICK WAS MELTED, BUT DID NOT GET HOT DURING TESTING IN THE LABORATORY. A REPLACEMENT POWER BRICK WAS USED TO POWER UP THE COMMUNICATOR AND THE UNIT PASSED ALL BASELINE TESTS. LABORATORY ANALYSIS CONFIRMED THE OBSERVATIONS OF A MELTED POWER BRICK.

Additional Manufacturer Narrative · 1

LATITUDE PATIENT SERVICES SENT THE PATIENT A RETURN LABEL AND REQUESTED THE PATIENT RETURN THE UNIT. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS REPORT WILL BE UPDATED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PATIENT REPORTED THEIR POWER BRICK MELTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COMMUNICATOR ACCESSORY LWS EXTERNAL MANUFACTURER 6482

Patients

Seq Age Sex Outcome Treatment
1 53 YR 0185| E030| T177