ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2011-05173
- Event Type
- Malfunction
- Date Received
- July 23, 2011
- Date of Event
- June 23, 2011
- Report Date
- June 24, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE, THE RESULTS WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT. NO CONCLUSIONS CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
(B)(4). DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP HISTORY SHOWS UNEXPLAINED REBOOTING EVENTS. VISUAL INSPECTION OF THE PUMP DID NOT REVEAL ANY DAMAGE. THE BATTERY CAP WAS ABLE TO SECURELY FASTEN TO THE PUMP. THE PUMP WAS ABLE TO BOOT UP PROPERLY AND WAS EXERCISED FOR 24 HOURS WITH NO ISSUES. THE PUMP WAS OPENED AND NO DAMAGE WAS OBSERVED.
THE PATIENT SAID THE PUMP IS REBOOTING WITHOUT USER INTERVENTION FOR THE PAST TWO DAYS. SHE SAID THE BATTERY CAP IS SECURE AND THAT THE YELLOW O-RING WAS NOT SHOWING. THERE WAS NO STRUCTURAL DAMAGE NOTED TO THE PUMP AND THE BATTERY CAP IS LESS THAN SIX MONTHS OLD. THERE WAS NO EVIDENCE OF MISUSE OF THE PRODUCT. THERE WAS NO VISIBLE CORROSION IN THE BATTERY COMPARTMENT. THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR |