MANUAL ORTHOPEDIC SURGICAL INSTRUMENT

MDR report key: 2173577 · Received July 23, 2011

Report

Report Number
1030489-2011-00920
Event Type
Malfunction
Date Received
July 23, 2011
Date of Event
June 22, 2011
Report Date
July 19, 2011
Manufacturer
MEDTRONIC SOFAMOR DANEK INSTRUMENT MANUFACTURING
Product Code
LXH
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DRIVER HAS BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. ANALYSIS RESULTS ARE NOT AVAILABLE AT THE TIME OF THIS REPORT. A FOLLOW-UP REPORT WILL BE SENT WHEN THE ANALYSIS IS COMPLETE. DEVICE HISTORY RECORDS FOR THESE LOTS WERE REVIEWED. NO DOCUMENTATION WAS FOUND THAT WOULD INDICATE A NON-CONFORMANCE TO SPECIFICATIONS.

Additional Manufacturer Narrative · 1

VISUALLY CONFIRMED APPROXIMATELY ~3MM OF THE INSTRUMENT TIP HAS BEEN BROKEN OFF, CONSISTENT WITH INTERFACE DURING USAGE. FRACTURE SURFACE ANALYSIS REVEALS A FAIRLY FLAT FRACTURE SURFACE AND CIRCULAR MATERIAL FLOW, CONSISTENT WITH TORSIONAL OVERLOAD.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE DURING SURGERY. THE TIP WAS RETRIEVED. NO PATIENT COMPLICATIONS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MANUAL ORTHOPEDIC SURGICAL INSTRUMENT APPLIANCE, FIXATION, SPINAL INTERLAMINAL LXH MEDTRONIC SOFAMOR DANEK INSTRUMENT MANUFACTURING NA 470811

Patients

Seq Age Sex Outcome Treatment
1 SCREW