MANUAL ORTHOPEDIC SURGICAL INSTRUMENT
Report
- Report Number
- 1030489-2011-00919
- Event Type
- Malfunction
- Date Received
- July 23, 2011
- Date of Event
- June 22, 2011
- Report Date
- July 19, 2011
- Manufacturer
- MEDTRONIC SOFAMOR DANEK INSTRUMENT MANUFACTURING
- Product Code
- LXH
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THE DRIVER HAS BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION. ANALYSIS RESULTS ARE NOT AVAILABLE AT THE TIME OF THIS REPORT. A FOLLOW-UP REPORT WILL BE SENT WHEN THE ANALYSIS IS COMPLETE. DEVICE HISTORY RECORDS FOR THESE LOTS WERE REVIEWED. NO DOCUMENTATION WAS FOUND THAT WOULD INDICATE A NON-CONFORMANCE TO SPECIFICATIONS.
IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.
VISUALLY CONFIRMED APPROXIMATELY ~3MM OF THE INSTRUMENT TIP HAS BEEN BROKEN OFF, CONSISTENT WITH INTERFACE DURING USAGE. FRACTURE SURFACE ANALYSIS REVEALS A FAIRLY FLAT FRACTURE SURFACE AND CIRCULAR MATERIAL FLOW, CONSISTENT WITH TORSIONAL OVERLOAD.
IT WAS REPORTED THAT THE TIP OF THE DRIVER BROKE DURING SURGERY. THE TIP WAS RETRIEVED. NO PATIENT COMPLICATIONS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MANUAL ORTHOPEDIC SURGICAL INSTRUMENT | APPLIANCE, FIXATION, SPINAL INTERLAMINAL | LXH | MEDTRONIC SOFAMOR DANEK INSTRUMENT MANUFACTURING | NA | M05J0232 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | SCREW |