IBOT PMD
Report
- Report Number
- 3014522447-2025-00002
- Event Type
- Injury
- Date Received
- March 31, 2025
- Date of Event
- March 14, 2025
- Report Date
- March 31, 2025
- Manufacturer
- MOBIUS MOBILITY LLC.
- Product Code
- IMK
- UDI-DI
- 00857584008010
- PMA / PMN Number
- K210920
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NH, US
- Reporter Occupation
- OTHER
Narratives
MOBIUS MOBILITY INTERVIEWED THE USER AND RETRIEVED THE EVENT/ALARM DATA AND BLACK-BOX DATA FROM THE DEVICE. THE DATA CONFIRMS THAT ON 3/12/25, AT 13:11:00, THE DEVICE ROLL LIMIT WAS EXCEEDED WHILE IN 4-WHEEL MODE, INDICATING FAILURE OF THE OPERATOR TO PROPERLY CONTROL THE DEVICE. BLACK BOX DATA SHOWS THE DEVICE MOVING FORWARD ON LEVEL GROUND IN 4-WHEEL MODE, THEN DECELERATING. THE CLUSTER ANGLE DIPS JUST PRIOR TO THE ROLL ANGLE INDICATING THE TIP. THE DATA IS CONSISTENT WITH THE USER DESCRIPTION OF NOT GOING OFF THE CURB STRAIGHT CAUSING THE DEVICE TO TIP OVER SIDEWAYS. NO DEVICE MALFUNCTION WAS INDICATED IN THE LOGS. ASSESSMENT OF THE LOGS CORROBORATED THE USER'S ACCOUNT THAT A FALL HAD OCCURRED AND INDICATED THAT THE FALL WAS DUE TO USER ERROR.
USER'S WIFE CALLED INDICATING HER HUSBAND HAD TIPPED OVER SIDEWAYS IN HIS IBOT WHILE AWAY ON VACATION. SHE SAID HE WAS GOING OFF A CURB WITH A GLASS IN HIS HAND AND TIPPED THE DEVICE OVER, BREAKING THE GLASS, WHICH RESULTED IN A SMALL CUT. THE CUT FROM THE GLASS RESULTED IN 4 STICHES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 912564 | IBOT PMD | IBOT | IMK | MOBIUS MOBILITY LLC. | IBOT PMD | N/A | 00857584008010 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR | Male | Required Intervention |