FDA Adverse Event Injury Summary report: N

IBOT PMD

MDR report key: 21734769 · Received March 31, 2025

Report

Report Number
3014522447-2025-00002
Event Type
Injury
Date Received
March 31, 2025
Date of Event
March 14, 2025
Report Date
March 31, 2025
Manufacturer
MOBIUS MOBILITY LLC.
Product Code
IMK
UDI-DI
00857584008010
PMA / PMN Number
K210920
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

MOBIUS MOBILITY INTERVIEWED THE USER AND RETRIEVED THE EVENT/ALARM DATA AND BLACK-BOX DATA FROM THE DEVICE. THE DATA CONFIRMS THAT ON 3/12/25, AT 13:11:00, THE DEVICE ROLL LIMIT WAS EXCEEDED WHILE IN 4-WHEEL MODE, INDICATING FAILURE OF THE OPERATOR TO PROPERLY CONTROL THE DEVICE. BLACK BOX DATA SHOWS THE DEVICE MOVING FORWARD ON LEVEL GROUND IN 4-WHEEL MODE, THEN DECELERATING. THE CLUSTER ANGLE DIPS JUST PRIOR TO THE ROLL ANGLE INDICATING THE TIP. THE DATA IS CONSISTENT WITH THE USER DESCRIPTION OF NOT GOING OFF THE CURB STRAIGHT CAUSING THE DEVICE TO TIP OVER SIDEWAYS. NO DEVICE MALFUNCTION WAS INDICATED IN THE LOGS. ASSESSMENT OF THE LOGS CORROBORATED THE USER'S ACCOUNT THAT A FALL HAD OCCURRED AND INDICATED THAT THE FALL WAS DUE TO USER ERROR.

Description of Event or Problem · 0

USER'S WIFE CALLED INDICATING HER HUSBAND HAD TIPPED OVER SIDEWAYS IN HIS IBOT WHILE AWAY ON VACATION. SHE SAID HE WAS GOING OFF A CURB WITH A GLASS IN HIS HAND AND TIPPED THE DEVICE OVER, BREAKING THE GLASS, WHICH RESULTED IN A SMALL CUT. THE CUT FROM THE GLASS RESULTED IN 4 STICHES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
912564 IBOT PMD IBOT IMK MOBIUS MOBILITY LLC. IBOT PMD N/A 00857584008010

Patients

Seq Age Sex Outcome Treatment
1 58 YR Male Required Intervention