FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2173445 · Received July 23, 2011

Report

Report Number
2531779-2011-05160
Event Type
Malfunction
Date Received
July 23, 2011
Report Date
June 27, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. PRODUCT ANALYSIS (PA) INVESTIGATED THE SUBJECT PUMP AND NOTED THE FOLLOWING : TESTING CONFIRMED INTERMITTENT RESPONSES TO BUTTON PRESSES ON THE OK BUTTON. THE PUMP WAS PLACED IN SUSPENSION AND THEN RESUME. THE OK BUTTON WORKED AFTER MULTIPLE PRESSES. PA CONFIRMED DAMAGE TO THE KEYPAD-THE KEYPAD IS TORN OVER THE OK BUTTON, EXPOSING THE BUTTON CONTACT. ALSO OBSERVED THE OK BUTTON CONTACT IS INVERTED. PA REMOVED KEYPAD COVER TO CHECK CONDITION OF THE BUTTON CONTACTS AND NOTICED CONTAMINATION WAS PRESENT UNDER THE CONTACTS OF ALL BUTTONS.

Description of Event or Problem · 1

THE PATIENT CLAIMED THAT THE OK BUTTONS DOES NOT RESPOND. THE KEYPAD IS REPORTEDLY INTACT. THERE WAS NO ADVERSE EVENT ASSOCIATE WITH THIS COMPLAINT. THE PUMP WAS REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1