FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2173433 · Received July 23, 2011

Report

Report Number
2531779-2011-05156
Event Type
Malfunction
Date Received
July 23, 2011
Report Date
June 27, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS N RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THE SUBJECT PUMP WAS INVESTIGATED WITH THE FOLLOWING FINDINGS: TESTING CONFIRMED INTERMITTENT RESPONSES TO BUTTON PRESSES ON THE KEYPAD. ALL THE KEYS INTERMITTENTLY STICK CAUSING SCROLLING. PRODUCT ANALYSIS (PA) OBSERVED DAMAGE TO THE KEYPAD. PA REMOVED KEYPAD COVER TO CHECK CONDITION OF THE BUTTON CONTACTS AND NOTICED THAT CONTAMINATION WAS PRESENT UNDER THE CONTACTS OF ALL BUTTONS.

Description of Event or Problem · 1

THE PATIENT CLAIMED THAT OK BUTTON IS NOT WORKING AND WHEN PRESSING THE OK BUTTON IT TAKES THE PATIENT TO ANOTHER SCREEN. THE KEYPAD IS REPORTEDLY INTACT. THERE WAS NO ADVERSE EVENT ASSOCIATE WITH THIS COMPLAINT. THE PUMP WAS REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 16 YR