MINI-AO TORQUE LIMITING DRIVER, 10INLB
Report
- Report Number
- 3025141-2025-00123
- Event Type
- Malfunction
- Date Received
- March 31, 2025
- Date of Event
- March 5, 2025
- Report Date
- March 31, 2025
- Manufacturer
- ACUMED, LLC
- Product Code
- LXH
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
MANUFACTURING AND INSPECTION RECORDS WERE REVIEWED, AND NO ANOMALIES WERE FOUND. THE RETURNED 1.5MM HEX DRIVER TIP, LOCKING GROOVE (80-0728) AND THE MINI-AO TORQUE LIMITING DRIVER (80-1008) WERE EXAMINED VISUALLY UNDER MAGNIFICATION. THE FRACTURE SURFACE OF THE DRIVER SPIRALS ALONG THE LENGTH OF THE TIP, ANGLING UPWARD TO A SHARP POINT, INDICATING POTENTIAL TORSIONAL LOADING TO FAILURE. THE FRACTURE SURFACE IS DULL AND GRAINY IN APPEARANCE, INDICATING A BRITTLE FRACTURE FROM A SINGLE OVERLOAD EVENT. THE DRIVER HANDLE SHOWED NO SIGNIFICANT OUTWARD SINGS OF WEAR OR DAMAGE. THE HANDLE CONNECTOR WAS UNSCREWED SLIGHTLY FROM THE MAIN BODY WITH A VISABLE GAP BETWEEN THE ORANGE HANDLE AND THE DRIVER CONNECTOR PORTION OF THE HANDLE. FUNCTIONAL TESTING CONFIRMED THAT THE DRIVER HANDLE WAS ABLE TO SUCCESSFULLY LOCK IN A DRIVER. HEX TIP BREAKAGE MAY OCCUR WHEN EXCESSIVE FORCE IS APPLIED TO THE DRIVER DURING USE TO OVERCOME INCREASED RESISTANCE. BASED ON THE INFORMATION RECEIVED AND DUE TO UNKNOWN SURGICAL CONDITIONS, THE ROOT CAUSE COULD NOT BE DETERMINED. RELATED SUBMISSION NAME: 3025141-2025-00122 AND 3025141-2025-00124.
WHILE IMPLANTING 2.3MM SCREWS IN THE DISTAL ROW, THE 1.5 MM HEX DRIVER TIP BROKE. THE ORANGE HANDLE ALSO MESSED UP. A DIFFERENT DRIVER TIP AND HANDLE WAS USED TO COMPLETE THE CASE. A DELAY OF 15 MINUTES WAS REPORTED AND THE PATIENT WAS STABLE WITH NO OTHER REPORTED EFFECTS. REPORT IS 2 OF 3.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 913520 | MINI-AO TORQUE LIMITING DRIVER, 10INLB | ORTHOPEDIC MANUAL SURGICAL INSTRUMENT | LXH | ACUMED, LLC | 80-1008 | 285214 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Female |