FDA Adverse Event Malfunction Summary report: N

INTERMATE

MDR report key: 2173347 · Received July 23, 2011

Report

Report Number
6000001-2011-13534
Event Type
Malfunction
Date Received
July 23, 2011
Date of Event
June 1, 2011
Report Date
June 27, 2011
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
FRN
PMA / PMN Number
K910425
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE FOR EVALUATION. THE REPORTED CONDITION OF A LEAK WAS NOT CONFIRMED. HOWEVER, VISUAL EXAMINATION OF THE SAMPLE NOTED A RUPTURED BLADDER IN A NON-FOOTED POSITION. AN ADDITIONAL COMPLAINT FILE WAS OPENED TO INVESTIGATE THE CONDITION FOUND IN SERVICE. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT. ADDITIONAL INFORMATION: PER REVIEW OF THE BATCH RECORDS, NO NONCONFORMANCE REPORT WAS DOCUMENTED FOR THIS LOT. ALL RELEASE CRITERIA WERE MET FOR THE BUILD OF THE LOT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL NARRATIVE: THE DEVICE IS AVAILABLE FOR EVALUATION PER THE CUSTOMER; HOWEVER, THE DEVICE HAS NOT YET BEEN RECEIVED BY BAXTER. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

BAXTER (B)(4) RECEIVED A REPORT THAT AN INTERMATE LEAKED AT AN UNSPECIFIED LOCATION. THE CONTENTS OF THE DEVICE ARE CURRENTLY UNKNOWN. THIS CONDITION HAS THE POTENTIAL TO INTERRUPT THERAPY OR BREACH THE STERILE FLUID PATHWAY. THE FACILITY DID NOT HAVE INFORMATION REGARDING WHETHER THERE HAVE BEEN ANY REPORTS OF PATIENT INJURY OR MEDICAL INTERVENTION IN ASSOCIATION WITH THIS EVENT. THE SAMPLE IS AVAILABLE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INTERMATE PUMP, INFUSION FRN BAXTER HEALTHCARE - IRVINE 10B075

Patients