FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2173334 · Received July 23, 2011

Report

Report Number
2531779-2011-05149
Event Type
Malfunction
Date Received
July 23, 2011
Report Date
June 25, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. THE SUBJECT PUMP WAS INVESTIGATED AND THE FOLLOWING WAS NOTED: THE "UP, DOWN, & OK" BUTTONS WERE NOT RESPONDING. MOISTURE RESIDUAL WAS LOCATED ON THE DISPLAY SCREEN. PRODUCT ANALYSIS PERFORMED LEAK TEST AND FOUND LEAKAGE FROM DISPLAY LENS. DISASSEMBLED THE PUMP AND FOUND MOISTURE RESIDUAL ON PCB AND SCREEN. REMOVED THE KEYPAD AND FOUND ADHESIVE AND CORROSION UNDER THE CONTACTS.

Description of Event or Problem · 1

THE PATIENT CLAIMED THAT THE BUTTONS ARE ALL UNRESPONSIVE. PATIENT STATED HE SWAM WITH PUMP THIS MORNING AND NOW REPORTS 2-3 DROPS OF WATER UNDER DISPLAY. THE KEYPAD IS REPORTEDLY INTACT. THERE WAS NO ADVERSE EVENT ASSOCIATE WITH THIS COMPLAINT. THE PUMP WAS REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 23 YR