FDA Adverse Event Injury Summary report: N

ALTRUA

MDR report key: 2173318 · Received July 23, 2011

Report

Report Number
2124215-2011-10782
Event Type
Injury
Date Received
July 23, 2011
Date of Event
June 10, 2011
Report Date
May 1, 2024
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NVZ
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 1

TO DATE, THE DEVICE AND LEAD REMAIN IMPLANTED. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED AND AN AMENDED REPORT SUBMITTED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

TO DATE THE EXPLANTED DEVICE HAS NOT BEEN RECEIVED AT BOSTON SCIENTIFIC. UPON RECEIPT THE DEVICE WILL UNDERGO DETAILED LABORATORY ANALYSIS IN AN ATTEMPT TO CONFIRM AND DETERMINE THE ROOT CAUSE OF THIS EVENT. AT THAT TIME A FINAL REPORT WILL BE SENT AFTER ANALYSIS IS COMPLETED.

Additional Manufacturer Narrative · 1

THE DEVICE AND LEAD HAVE BEEN EXPLANTED. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

APPROXIMATELY (B)(6) MONTHS LATER, THIS PACEMAKER WAS NOTED TO BE IN UNIPOLAR MODE AS A LEAD SAFETY SWITCH HAD OCCURRED DUE TO OUT OF RANGE IMPEDANCE MEASUREMENTS ON THE RIGHT VENTRICULAR (RV) LEAD GREATER THAN 2500 OHMS. IN ADDITION, A HOLTER MONITOR REPORT WAS REVIEWED AND REVEALED HIGH ATRIAL RATES FOLLOWED BY EXTENDED PAUSES GREATER THAN 7 SECONDS WITH NO PACEMAKER STIMULATION. THE PATIENT WAS NOT SYNCOPAL DURING THESE EPISODES, HOWEVER, FELT DIZZY SPELLS. MOST SYMPTOMS OCCUR WHILE SHE IS DRIVING OR LYING ON HER RIGHT SIDE. A LEAD AND HEADER CONNECTION ISSUE WAS SUSPECTED. A REVISION PROCEDURE WAS PERFORMED, AND THE LEADS WERE DISCONNECTED FROM THE HEADER AND INSERTED INTO A NEW DEVICE. THE LEADS WERE INSERTED INTO A PACING SYSTEM ANALYZER (PSA) AND TESTED AND REVEALED ALL OTHER MEASUREMENTS WERE STABLE AND WITHIN NORMAL RANGE VISUAL INSPECTION OF THE DEVICE REVEALED DAMAGE TO THE OUTSIDE OF THE CAN AS MANIPULATION WAS USED TO REMOVE THE DEVICE. ALSO VISUAL INSPECTION OF THE DEVICE AND LEAD REVEALED NO HEADER OR CONNECTION ISSUE. A BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) AGENT WAS CONTACTED ON DEVICE PROGRAMMING FOR OPTIMIZATION FOR PATIENT. THE DEVICE WAS EXPLANTED AND SENT BACK FOR ANALYSIS. THE RIGHT VENTRICULAR (RV) LEAD REMAINS IN SERVICE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED THAT DURING A RECENT DEVICE CHANGE OUT ON (B)(6)2012, THIS RIGHT VENTRICULAR (RV) LEAD WAS EXPLANTED DUE TO A SUSPECTED LEAD ISSUE. NO ADVERSE PATIENT EFFECTS WERE REPORTED. TO DATE IT, THE LOCATION OF THE LEAD IS UNKNOWN.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING A ROUTINE FOLLOW-UP VISIT, THIS PACEMAKER WAS NOTED TO BE IN UNIPOLAR MODE AS A LEAD SAFETY SWITCH HAD OCCURED DUE TO OUT OF RANGE IMPEDANCE MEASUREMENTS ON THE RIGHT VENTRICULAR (RV) LEAD. IN ADDITION, A HOLTER MONITOR REPORT WAS REVIEWED AND REVEALED HIGH ATRIAL RATES FOLLOWED BY EXTENDED PAUSES WITH NO PACEMAKER STIMULATION. THE PATIENT WAS NOT SYNCOPAL DURING THESE EPISODES, HOWEVER, FELT 'HEAD RUSHES', SHORTNESS OF BREATH, AND PALPITATIONS. DURING DEVICE TESTING, NOISE AND EXTENDED PACING PAUSES OF THREE SECONDS WERE NOTED ON STORED ELECTROGRAMS AND WHEN PRESSURE WAS APPLIED TO THE DEVICE POCKET. A CONNECTION OR HEADER ISSUE WAS SUSPECTED. THE DEVICE WAS REPROGRAMMED FROM A SINGLE CHAMBER INHIBITED RATE RESPONSIVE (SSIR) MODE TO A SINGLE CHAMBER FIXED PACING RATE RESPONSIVE (SOOR) MODE AND THE PACING VOLTAGE WAS INCREASED FROM 2 TO 3V. A CHEST X-RAY WAS PERFORMED AND REVEALED NO EVIDENCE OF HEADER DISPLACEMENT AND THE TIP OF THE LEAD PIN APPEARS TO EXTEND BEYOND THE APPROPRIATE, PROXIMAL ELECTRICAL CONNECTOR BLOCK IN THE HEADER OF THE DEVICE. THEREFORE, THERE WAS NO OBVIOUS EXPLANATION FOR THE MALFUNCTION OF THE PACING SYSTEM. THE PATIENT HAS HAD NO FURTHER SEVERE HEAD RUSHES SINCE THE RECENT DEVICE REPROGRAMMING; HOWEVER, SHE HAS FELT HER HEART BEATING VERY FORCEFULLY DURING THE NIGHT, LIKELY RELATED TO THE HIGHER STIMULATION VOLTAGE THAT WAS PROGRAMMED. THE POTENTIAL NEED FOR REPEAT SURGERY HAS BEEN DISCUSSED WITH THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ALTRUA IMPLANTABLE PULSE GENERATOR NVZ GUIDANT CRM CLONMEL IRELAND S501

Patients

Seq Age Sex Outcome Treatment
1 58 YR Female Life Threatening| O 4470.| S501.| (B)(4)| (B)(4)