CONTAK RENEWAL
Report
- Report Number
- 2124215-2011-11175
- Event Type
- Injury
- Date Received
- July 23, 2011
- Date of Event
- June 15, 2011
- Report Date
- August 1, 2011
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NIK
- PMA / PMN Number
- P010012
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED. THE DEVICE MEMORY CONFIRMED AN OUT OF RANGE IMPEDANCE MEASUREMENT FOR THE RIGHT VENTRICULAR CHANNEL.
NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS EVENT WILL BE UPDATED.
THE PRODUCT HAS NOT BEEN RETURNED. IF RETURNED, OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS EVENT WILL UPDATED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE RIGHT VENTRICULAR LEAD EXHIBITED HIGH SHOCK IMPEDANCE MEASUREMENTS. ADDITIONALLY, THERE WAS A RED ALERT AS THE DEVICE HAD BEEN PROGRAMMED OFF DUE TO A LEAD FRACTURE AND IMPENDING REVISION PROCEDURE. NO ADVERSE PATIENT EFFECTS WERE NOTED.
INFORMATION WAS LATER RECEIVED THAT THIS DEVICE WAS EXPLANTED DUE TO A SYSTEM UPGRADE. NO ADVERSE PATIENT EFFECTS WERE NOTED AS A RESULT OF THE PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONTAK RENEWAL | IMPLANTABLE CHF GENERATOR | NIK | GUIDANT CRM CLONMEL IRELAND | H210 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR | 4136| 4538| H210| 0185 |