FDA Adverse Event Injury Summary report: N

CONTAK RENEWAL

MDR report key: 2173286 · Received July 23, 2011

Report

Report Number
2124215-2011-11175
Event Type
Injury
Date Received
July 23, 2011
Date of Event
June 15, 2011
Report Date
August 1, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
PMA / PMN Number
P010012
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED. THE DEVICE MEMORY CONFIRMED AN OUT OF RANGE IMPEDANCE MEASUREMENT FOR THE RIGHT VENTRICULAR CHANNEL.

Additional Manufacturer Narrative · 1

NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS EVENT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

THE PRODUCT HAS NOT BEEN RETURNED. IF RETURNED, OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS EVENT WILL UPDATED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE RIGHT VENTRICULAR LEAD EXHIBITED HIGH SHOCK IMPEDANCE MEASUREMENTS. ADDITIONALLY, THERE WAS A RED ALERT AS THE DEVICE HAD BEEN PROGRAMMED OFF DUE TO A LEAD FRACTURE AND IMPENDING REVISION PROCEDURE. NO ADVERSE PATIENT EFFECTS WERE NOTED.

Description of Event or Problem · 1

INFORMATION WAS LATER RECEIVED THAT THIS DEVICE WAS EXPLANTED DUE TO A SYSTEM UPGRADE. NO ADVERSE PATIENT EFFECTS WERE NOTED AS A RESULT OF THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTAK RENEWAL IMPLANTABLE CHF GENERATOR NIK GUIDANT CRM CLONMEL IRELAND H210

Patients

Seq Age Sex Outcome Treatment
1 78 YR 4136| 4538| H210| 0185