FDA Adverse Event Malfunction Summary report: N

CONTAK RENEWAL

MDR report key: 2173276 · Received July 23, 2011

Report

Report Number
2124215-2011-12198
Event Type
Malfunction
Date Received
July 23, 2011
Date of Event
May 3, 2011
Report Date
May 12, 2011
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, ANALYSIS WAS PERFORMED AT OUR POST MARKET QUALITY ASSURANCE LABORATORY. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THAT THIS DEVICE HAD A COMPROMISED LOW-VOLTAGE CAPACITOR THAT DID NOT MEET DESIGN SPECIFICATION. DESPITE THIS COMPROMISED CAPACITOR, CURRENT LEAKAGE WAS NOT SUFFICIENT TO ADVERSELY IMPACT DEVICE LONGEVITY. NORMAL PACING, SENSING, AND DEFIBRILLATION THERAPY FUNCTIONS WERE VERIFIED DURING TESTING.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE DID NOT PASS A PORTION OF THE RETURNED PRODUCTS TESTING, SUGGESTING A POSSIBLE PERFORMANCE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTAK RENEWAL IMPLANTABLE CHF GENERATOR NIK GUIDANT CRM CLONMEL IRELAND H170

Patients

Seq Age Sex Outcome Treatment
1 70 YR 1581| H170| MISMATCH