FDA Adverse Event Injury Summary report: N

FINELINE II

MDR report key: 2173169 · Received July 23, 2011

Report

Report Number
2124215-2011-10448
Event Type
Injury
Date Received
July 23, 2011
Date of Event
June 14, 2011
Report Date
June 14, 2011
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

AS OF TODAY, NO ADDITIONAL INFORMATION IS AVAILABLE AND OUR INVESTIGATION IS COMPLETE AT THIS TIME. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS REPORT WILL BE UPDATED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR (RV) LEAD DISPLAYED AN INCREASE IN IMPEDANCES AND PACING THRESHOLD. A LEAD FRACTURE WAS SUSPECTED. A LEAD REVISION PROCEDURE WAS PERFORMED AND THE LEAD WAS CUT AND CAPPED. A NEW RV LEAD WAS PLACED WITHOUT INCIDENT. THERE WERE NO ADDITIONAL ADVERSE PATIENT EFFECTS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FINELINE II IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4456

Patients

Seq Age Sex Outcome Treatment
1 79 YR Life Threatening| R 4456| S603| 4135