FDA Adverse Event Injury Summary report: N

ACUITY

MDR report key: 2173148 · Received July 23, 2011

Report

Report Number
2124215-2011-10412
Event Type
Injury
Date Received
July 23, 2011
Date of Event
June 13, 2011
Report Date
July 5, 2011
Manufacturer
CPI - DEL CARIBE
Product Code
OJX
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ATTEMPTS TO RETRIEVE ADDITIONAL INFORMATION HAVE BEEN MADE. AT THIS TIME THERE IS NO ADDITIONAL INFORMATION AVAILABLE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

ATTEMPTS TO RETRIEVE ADDITIONAL INFORMATION HAVE BEEN UNSUCCESSFUL. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION RECEIVED FROM THE LOCAL AREA SALES REPRESENTATIVE INDICATED THE LV LEAD WAS EXPLANTED DURING A DEVICE CHANGEOUT PROCEDURE. IT WAS REPORTED THE LEAD HAD A CONDUCTOR FRACTURE. IT WAS DISCARDED BY THE HOSPITAL STAFF AND WILL NOT BE RETURNED FOR ANALYSIS. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE LEFT VENTRICULAR (LV) LEAD WAS EXHIBITING HIGH PACING IMPEDANCE MEASUREMENTS GREATER THAN 2,500 OHMS. THE PATIENT WAS PACEMAKER DEPENDENT AND LV CAPTURE WAS UNABLE TO BE OBTAINED. TO DATE, THERE HAVE BEEN NO ADVERSE PATIENT EFFECTS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACUITY IMPLANTABLE LEAD OJX CPI - DEL CARIBE 4554

Patients

Seq Age Sex Outcome Treatment
1 77 YR Hospitalization| L| R H120| 4554| 4087| 4086