FDA Adverse Event Malfunction Summary report: N

ENDOTAK RELIANCE

MDR report key: 2173108 · Received July 23, 2011

Report

Report Number
2124215-2011-10483
Event Type
Malfunction
Date Received
July 23, 2011
Date of Event
June 13, 2011
Report Date
June 13, 2011
Manufacturer
CPI - DEL CARIBE
Product Code
NVY
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

AT THIS TIME THE PRODUCT REMAINS IN SERVICE. IF ADDITIONAL INFORMATION BECOMES AVAILABLE THIS REPORT WILL BE UPDATED AS NECESSARY.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEAD EXHIBITED NOISE AS WELL AS VARIED IMPEDANCE MEASUREMENTS GOING HIGH AS WELL AS LOW. ONCE SOME HOSPITAL EQUIPMENT WAS TURNED OFF IN THE ROOM, ALL MEASUREMENTS RETURNED TO NORMAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOTAK RELIANCE IMPLANTABLE LEAD NVY CPI - DEL CARIBE 0184

Patients

Seq Age Sex Outcome Treatment
1 64 YR 4517| N118| 0184| 4469| H175